Laboratory Centrifuge Operation and Maintenance Procedures

Release time:

2013-01-21


Overview:

During ultracentrifugation, the liquid must completely fill the centrifuge tubes. If vacuum is required during ultracentrifugation, only by filling the tubes completely can tube deformation be avoided. If the centrifuge tube caps have poor sealing, the tubes should not be filled to capacity to prevent leakage, which could interfere with the proper functioning of the sensor.

In practical operation, the handling and maintenance of laboratory centrifuges should be standardized to ensure proper instrument performance and also to safeguard the safety of personnel.

  1. Turn on the centrifuge switch and enter standby mode.

  2. Select the appropriate rotor: During centrifugation, the volume of liquid in each centrifuge tube must not exceed two-thirds of the total capacity; otherwise, the liquid is likely to...

  For overflow: Before and after use, pay attention to whether there is any residual liquid leakage inside the rotor; ensure it remains dry. Switch over.

  When turning the rotor, make sure to securely lock the spindle of the centrifuge and the rotor’s latch.

  2. The balance error of centrifuge tubes should be within 0.1 gram.

  3. Select centrifugation parameters:

  a) Press the temperature setting button, then use the numeric keys to set the centrifugation temperature, and press Enter to confirm.

  b) Press the speed setting button to switch between RPM/RCF setting modes, and use the numeric keys to set the centrifugal speed.

  Press Enter to confirm.

  c) Press the rotor setting button, then use the numeric keys to set the rotor model, and press Enter to confirm.

  d) Press the time-setting button, then use the numeric keys to set the centrifugation time, and press Enter to confirm.

  e) The braking or acceleration rate of the centrifuge is generally set between 0 and 4 and should not be adjusted frequently.

  4. Place the balanced centrifuge tubes symmetrically into the rotor. Close the rotor lid and tighten the screws.

  5. Press the centrifuge lid door. If the lid door is not securely closed, the centrifuge will not start.

  6. Press the START button to begin centrifugation. After centrifugation starts, wait until the centrifugal speed reaches the set speed before proceeding.

  Turn on the centrifuge, and once you detect any abnormalities (such as imbalance causing noticeable vibration or excessively loud noise),

  Immediately press the STOP button, and if necessary, directly turn off the power by pressing the power switch to stop the centrifuge and identify the cause.

  7. If the machine malfunctions, please contact the relevant personnel promptly.

  8. After use, please clean the rotor and the centrifuge chamber. Do not close the centrifuge lid to allow moisture to evaporate.

  9. After use, you must register and specify the details of how it was used.

  Precautions

  1. When the centrifuge is in the pre-cooling state, the centrifuge lid must be closed. After the centrifugation is complete, remove the rotor and place it upside down on the lab bench. Wipe dry any remaining water inside the chamber, and leave the centrifuge lid open.

  2. When performing ultracentrifugation, the liquid must completely fill the centrifuge tube. If vacuum is required during ultracentrifugation, only a fully filled tube can prevent deformation of the tube. If the centrifuge tube cap has poor sealing, the liquid cannot be filled to capacity to avoid leakage, which could interfere with the proper functioning of the sensor.

  3. During precooling, the rotor cover can be placed on the centrifuge’s platform or on the lab bench. Under no circumstances should it be left floating loosely on the rotor without being securely tightened, because if the centrifuge is accidentally started, the rotor cover could fly off and cause an accident!

  4. After tightening the rotor cover, be sure to use your fingers to check that there is no gap between the rotor and the rotor cover. If there is a gap, unscrew and retighten the cover until you confirm that there is no gap before starting the centrifuge.

  5. Be sure to ground the device when using it. The substances added into the centrifuge tubes should be relatively balanced. If an imbalance occurs between the two sides, it could cause significant damage to the centrifuge and, at the very least, shorten its service life.

  6. During the centrifugation process, operators must not leave the centrifuge room. If any abnormal situation occurs, operators must not turn off the power supply (POWER); instead, they should press the STOP button. Before pre-cooling, operators must complete the centrifuge usage log.