Operating and Usage Procedure for High-Speed Refrigerated Centrifuges

Release time:

2013-01-22


Overview:

For example, the TGL-20M high-speed refrigerated centrifuge has a maximum rotor speed of 20,000 rpm; however, if the centrifuge is too old, the speed must be reduced accordingly. Try not to set the rotor speed or time too high.

1. Set the operating temperature, typically 4℃. First, place the centrifuge rotor into the centrifuge chamber, making sure it is securely seated on the spindle. Close the chamber door; the temperature of the centrifuge chamber will then begin to drop. Allow sufficient time for precooling.

2. Place the samples into appropriate centrifuge tubes, balance their weights using a pair of scales, and then securely close and tighten the lids on the centrifuge tubes.

Note that the centrifuge tubes should be filled only to 70% capacity. Even though they are equipped with lids, they may still leak due to excessive centrifugal force.
Most centrifuge tubes come with lids; be sure to balance the lids along with the tubes themselves.
Centrifuge tubes are usually placed on crushed ice. When removing them for balancing, be sure to wipe off any liquid from the crushed ice thoroughly.
The single centrifuge tube must be balanced by another centrifuge tube filled with clean water.

3. Place the balanced centrifuge tubes symmetrically into the rotor, then close the rotor lid, making sure it is securely tightened.

If the lid of the centrifuge top isn't tightened properly, it could fall off during centrifugation, causing serious injury!

4. Close the centrifuge chamber door and set the desired speed and time on the control panel—for example: 8,000 rpm, 30 minutes.

Please note that the rotational speed used must never exceed the centrifuge’s specified limit. The maximum rotational speed is usually indicated on the centrifuge itself—for example, the TGL-20M high-speed refrigerated centrifuge has a maximum speed of 20,000 rpm. However, if the centrifuge is too old, you must reduce the speed accordingly. As a general guideline, avoid setting the rotational speed and time too high—for instance, if 7,000 rpm is sufficient, there’s no need to set it at 8,000 rpm.

5. After confirming that all steps are correct, press the Start button to begin operation. The centrifuge will gradually accelerate; at this point, closely monitor its performance.

For some older time-setting knobs, when setting a duration of less than 5 minutes, you need to first turn the knob past 10 minutes and then turn it back to 5 minutes.
When you press the Start button, don't release it too quickly—hold it down for about one second before letting go.
It’s most dangerous when you start accelerating. If you notice any unusual sounds or significant vibrations, immediately press Stop.

6. After the centrifuge reaches the desired speed, make sure everything is normal before leaving.

7. After completing the centrifugation, wait until the rotor has come to a complete stop before opening the chamber door. Please remove the centrifuge tubes as soon as possible, first checking whether the tubes are intact and noting the position of the pellet. Quickly decant the supernatant, taking care not to disturb or cloud the pellet.

Whether you need the supernatant or the precipitate, please use a clean beaker to collect the supernatant to avoid any mistakes.
Be very careful when pouring off the supernatant to avoid dumping the precipitate along with it; if you accidentally mix them together, you’ll have to start over.
If there is a leak in the centrifuge tube, identify the cause and immediately clean the rotor or the centrifuge chamber.

8. During the interval between two centrifugation cycles, there is no need to remove the centrifuge rotor; however, keep the cabin door closed to prevent hot air from entering the centrifuge chamber.

9. After you’ve finished using the centrifuge, remove the rotor and clean it. You can rinse it with tap water and then let it air-dry upside down. Be sure to keep the rotor chamber door open. Once the ice has melted, give it a light wipe-down and thorough cleaning. If any liquid has leaked out, immediately wash it off with clean water. Also, take a moment to check and clean the balance used for balancing the centrifuge tubes as well as the work surface—these areas tend to get dirty easily, so be sure to clean them carefully before leaving.