Causes of Imbalance in Medical Centrifuge Rotors

Release time:

2013-01-17


Overview:

During sample loading into medical centrifuge tubes, if the tubes are overfilled with sample liquid or the tube caps are not tightly screwed shut, the high-vacuum environment within the centrifuge chamber during centrifugation can cause the tubes to flatten, rupture, and leak sample liquid. This imbalance can lead to rotor instability, resulting in shaft bending or even shaft breakage accidents.

When we use laboratory centrifuges, it is common for the centrifuge rotor to fail to operate properly due to imbalance, or for accidents such as shaft breakage to occur when the rotor operates in an unbalanced state. We have summarized below the common causes of rotor imbalance in commonly used medical centrifuges.

1. During balancing of medical centrifuge tubes, due to prolonged use of the balance without calibration, the asymmetry of symmetrically placed centrifuge tubes exceeded 0.1 to 1 gram.

2. During sample loading into medical centrifuge tubes, if the tubes are overfilled with sample liquid or the tube caps are not tightly screwed shut, the high-vacuum environment within the centrifuge chamber during centrifugation can cause the tubes to collapse, rupture, and result in leakage of the sample liquid. This imbalance can lead to rotor misalignment, shaft bending, or even shaft breakage.

3. An accident occurred because the specific gravity of the original sample was not equal to that of the balancing liquid, causing an imbalance in the rotor's dynamic balance.

4. When using a horizontal rotor, failing to carefully match the serial number of the bucket with the serial number of the rotor body can affect the rotor's dynamic balance.

5. Accidents occur when aluminum alloy centrifuge tube caps and stainless steel centrifuge tube caps are used interchangeably (due to their different specific gravities).

6. Improper use and inadequate sterilization of the rubber sealing rings inside the centrifuge tube caps and rotor cover—such as prolonged use leading to internal fractures, or exposure to high-temperature sterilization (e.g., baking in a drying oven)—can cause these rings to age, crack, and lose their sealing effectiveness. As a result, samples may spill out during high-speed operation, causing the rotor to run in an unbalanced state.

7. For centrifuge tubes made of various materials, failure to thoroughly understand the manufacturer’s specified usage limits and disinfection methods before use led to improper handling with chemical solvents and inappropriate disinfectants, causing the centrifuge tubes to swell and rupture during operation, resulting in accidents.

8. Due to carelessness during work, the rotor cover was not tightened properly or the rotor cover was mistakenly swapped with the rotor shaft, resulting in mismatched screw threads. When the machine was started and operated, the rotor cover flew off, causing a severe shaft breakage accident.

Please ensure that the centrifuge rotor is correctly and evenly positioned on the centrifuge spindle to prevent accidents.