Laboratory centrifuge calibration is both fundamental and critical.
Release time:
2021-05-10
Overview:
Laboratory centrifuge It has a wide range of applications and is commonly used in hospital laboratories, research institution laboratories, food and drug testing laboratories, food hygiene laboratories, and various other types of laboratories. Its rotational speed does not exceed 30,000 rpm.
In recent years, China’s laboratory centrifuge industry has developed rapidly and has now reached a considerable scale. Among them, centrifuges operating at speeds below 40,000 rpm have already attained or are nearing world-class standards. Starting from the first-generation laboratory centrifuges that relied on direct plug-in power to drive rotation for separation, after more than 30 years of relentless efforts and internal growth, the industry has made a quantum leap—from small-capacity to large-capacity models, from low-speed to high-speed and even ultra-high-speed units. More importantly, it has successfully adopted AC variable-frequency speed control, multi-functional automation, and microcomputer-based control systems, marking a transformative breakthrough.
The rotational speed of a laboratory centrifuge is a critically important technical parameter. The magnitude of the centrifugal force is directly proportional to the rotational speed of the centrifuge. If the rotational speed fails to meet the specified requirements, the suspension or emulsion cannot be completely separated. In the course of laboratory analysis, using different laboratory centrifuges to separate the same substance may lead to divergent conclusions, seriously affecting the accuracy of the analytical results.
(1) There has been significant progress in the variety, specifications, and capacity of centrifuges, with a virtually complete range of models and adapters available. These products now largely meet the needs of the market and the requirements of a broad customer base.
(2) The structure and control systems of machinery have made significant progress. As times have changed, people’s awareness has also undergone a noticeable shift. Today, people no longer focus solely on aspects such as machine structural strength, service life, and reliability; instead, they adopt a comprehensive design approach that takes into account mechanical safety, electrical controls, and other factors. Moreover, there is now a thorough consideration and design aimed at the machine’s environmental protection and aesthetic appearance, making the products more user-friendly and human-centered.
Since there are no verification procedures or calibration standards for laboratory centrifuge rotational speeds, it is impossible to carry out verification or calibration of centrifuges. Therefore, it is necessary to explore methods for calibrating the rotational speeds of laboratory centrifuges.
Environmental conditions
Ambient temperature: (20 ± 5)℃; relative humidity: ≤80%; no direct exposure to sunlight or other heat sources; no strong vibrations or air currents in the surrounding area; pollution level: Class 2; no conductive dust, explosive gases, high-concentration dust, or corrosive substances present in the surrounding environment.
Measurement standards
According to the metrological performance requirements of laboratory centrifuges, the rotational speed meter used must have an accuracy of 0.2% or better. Its expanded uncertainty U(k=2) shall meet the calibration requirements for the technical specifications of the laboratory centrifuge being measured (i.e., the expanded uncertainty value shall not exceed one-third of the absolute value of the measurement error).
Calibration method
1. Calibration of rotational speed error
Before calibration, the measurement equipment should be allowed to fully stabilize under the specified conditions. Set the laboratory centrifuge—equipped with a rotor loaded to its maximum capacity—to its highest rated speed and allow it to stabilize for 5 minutes. Use a tachometer to measure the centrifuge’s rotational speed, taking one reading every minute for a total of 10 readings. Calculate the speed error for each reading according to Formula (1), and take the maximum value obtained as the final result.
Rotational speed error calculation:

In the formula: δci —— the error of the i-th measurement, %; ni —— the i-th measured rotational speed, r/min; nmax —— the maximum rated rotational speed, r/min; i —— the measurement sequence number, with values ranging from 1, 2, 3… to 10.
2. Rotational Speed Stability Calibration
Set the laboratory centrifuge, loaded to its maximum capacity, to its highest rated speed and allow it to stabilize for 5 minutes. Use a tachometer to measure the centrifuge’s rotational speed, taking one measurement every minute for a total of 10 measurements (alternatively, you may use data recorded from measuring rotational speed errors). Calculate the arithmetic mean of the 10 measured speeds, and then compute the stability of each measurement using Equation (2). Take the maximum value obtained as the final result.
Rotational Speed Stability Calculation

In the formula: δdi —— the i-th instability, %; ni —— the measured rotational speed for the i-th trial, r/min; —— the average rotational speed, r/min; i —— the measurement serial number, with values ranging from 1, 2, 3… to 10.