On the Service Life of Centrifuges
Release time:
2015-09-17
Overview:
Centrifuge It is a high-speed rotating instrument that utilizes centrifugal force. A specialized piece of equipment that utilizes the principle of differential sedimentation rates of solids and liquids in a centrifugal field to rapidly separate and extract components from a mixed liquid.
Since most of the samples subjected to centrifugation are corrosive, even with isolation measures such as centrifuge tubes, some leakage is inevitable. Moreover, in the case of refrigerated centrifuges, the centrifuge chamber often remains frozen and humid during operation—a frequency of use that is easy to imagine for pharmaceutical manufacturers. We require that after each use, the centrifuge cups, rotor components, and centrifuge chamber be thoroughly wiped clean, dried thoroughly, and kept completely dry. However, most users fail to follow these instructions, which can lead to a gradual accumulation of potential hazards.
Centrifuge During the centrifugation process, under the combined effects of stress corrosion and chemical corrosion, the rotor will experience dual corrosion and fatigue. This can lead to cracks, deformation, and other forms of damage when the rotor is spinning at high speeds. In severe cases, it may even result in rotor rupture, posing serious safety hazards such as injury or harm to people. Therefore, once a centrifuge reaches its end-of-life, never take the cheap route out of economic considerations—doing so could lead to a major disaster, which would be detrimental both to yourself and to the organization.
For a centrifuge, the most critical component is the high-speed rotating rotor. According to industry standards, the service life of an aluminum alloy rotor is specified as 6 years. Under normal operating conditions, the cumulative running time per day is 5 hours (with additional auxiliary time for balancing, sample extraction, and other tasks). Therefore, the rotor's service life is 6 years. However, if the daily usage exceeds or continuously exceeds 5 hours, the rotor's service life must be evaluated based on actual conditions—never take this lightly.
Based on the above explanation, our company hereby solemnly declares that the use of products exceeding their service life is strictly prohibited. Centrifuge 。
Shanghai Luxiangyi Centrifuge Instrument Co., Ltd.
2015.9.17