How to Properly Install a Blood Bank Centrifuge
Release time:
2016-07-07
Overview:
How to install correctly Blood bank centrifuge
The development of centrifuge technology in China began after the founding of the People’s Republic of China. The first centrifuge with practical value was invented in 1954. Following the reform and opening-up policy, China began to vigorously introduce foreign centrifuge technologies. During the Seventh Five-Year Plan, centrifuges were designated as a key project for China’s scientific and technological research. After years of persistent efforts, China’s centrifuge industry has now reached a mature scale and has developed strong capabilities for independent R&D and innovation. Judging from current trends, the industry’s development prospects are very promising. However, compared with advanced centrifuge equipment available abroad, China’s centrifuge technology still has considerable room for improvement.
The blood bank centrifuge is specially designed by our company to meet the needs of rapid centrifugation in blood banks. This machine features microcomputer control and is driven by a brushless DC motor, ensuring stable operation, low noise, and high rotational speed accuracy. It is equipped with a touch panel for programmable operation, and its main operating parameters can be set according to requirements and automatically saved. In addition, it boasts a digital display (LCD screen) with a user-friendly interface, making operation simple and convenient.
Brief description of the blood bank centrifuge installation method:
1. Power Supply Voltage: Voltage fluctuations must comply with the national standard within a tolerance of ±10%. If voltage fluctuations exceed this range, it is recommended to use a voltage regulator to ensure compliance with the power grid requirements.
2. The grounding wire must be secure: The wiring of the room’s power supply must meet the required standards. In China, the power system uses a three-phase four-wire configuration. The three phases refer to the three-phase electricity used in industrial applications. When using three-phase electricity, the neutral wire and the ground wire must be kept separate. More importantly, whether it’s three-phase industrial power or ordinary single-phase power, the ground wire must be reliably connected to prevent electric leakage. Under no circumstances should the ground wire be connected to heating pipes or water pipes.
3. The floor must be level and stable: For large-scale blood bank centrifuges, the floor at the installation site must be firmly secured to ensure that the centrifuge rotor operates smoothly without sinking or deforming. The floor surface at the installation and fixing location should be both level and sturdy.
4. Adjust the level: Use a small, relatively accurate spirit level. With the blood bank centrifuge lid open, locate the horizontal position on the main shaft. Adjust the level by turning the adjusting bolts located next to the four casters. Be sure not to let the "L" foot go empty.
Keywords: Blood bank centrifuge