Determination of Antibiotic Residues in Aquatic Products by Liquid Chromatography
Release time:
2015-03-09
Overview:
1. Instruments and Reagents
HP1100 high-performance liquid chromatograph, G1315A DAD diode-array detector, G1313A ALS autosampler; SUPELCO DISCOVERY C18 5 μm, 250 × 4.6 mm chromatographic column. Solid-phase extraction cartridges are ZORBAX SPE C18 cartridges. 12-channel solid-phase extraction vacuum enrichment apparatus (Alltech Scientific Limited). Ultrasonic cleaner KS-500D, high-speed centrifuge GL-25M, meat grinder, vortex mixer. Methanol (chromatography grade), sodium dihydrogen phosphate (analytical grade), EDTA (analytical grade), perchloric acid solution (superior grade), nitric acid (superior grade). Standard substances of oxytetracycline, tetracycline, chlortetracycline, and chloramphenicol (SIGMA Company).
2. High-performance liquid chromatography analysis conditions
The chromatographic column was a SUPELCO DISCOVERY C18 5 μm, 250 × 4.6 mm. The mobile phase consisted of 0.01 mol/L sodium dihydrogen phosphate plus 0.001 mol/L EDTA: methanol at a ratio of 60:40 (v/v, pH = 2.0). The flow rate was 0.8 ml/min, the detection wavelength was 268 nm, the column temperature was 25℃, and the injection volume was 20 μl.
3. Experimental Method
(1) Preparation of standard stock solutions of oxytetracycline, tetracycline, chlortetracycline, and chloramphenicol
Weigh out 0.0100 mg each of oxytetracycline, tetracycline, chlortetracycline, and chloramphenicol standard substances separately and place them into four 10 ml volumetric flasks. Add methanol to each flask to dissolve the substances completely, then dilute to the mark with methanol. The resulting solutions have concentrations of 1.00 mg/ml each. Store these solutions in the refrigerator.
(2) Preparation of the mixed standard intermediate solution
Take 1.00 ml each of the oxytetracycline and tetracycline standard stock solutions, as well as 2.00 ml each of the chlortetracycline and chloramphenicol standard stock solutions. Transfer these aliquots into a 100 ml volumetric flask, dilute to the mark with mobile phase A, and obtain solutions with the following concentrations: oxytetracycline (10 μg/ml), tetracycline (10 μg/ml), chlortetracycline (20 μg/ml), and chloramphenicol (20 μg/ml). Store these solutions in the refrigerator.
(3) Preparation of standard stock solutions
Accurately pipette 0.1, 0.2, 0.5, 1.0, 2.0, 4.0, 6.0, 8.0, and 10.0 ml of the mixed standard intermediate solution into 10 ml volumetric flasks, then dilute to the mark with mobile phase A to prepare a series of standard solutions with concentrations of: 0.1, 0.2, 0.5, 1.0, 2.0, 4.0, 6.0, 8.0, and 10.0 μg/mL (for oxytetracycline and tetracycline); and 0.2, 0.4, 1.0, 2.0, 4.0, 8.0, 12.0, 16.0, and 20.0 μg/mL (for chlortetracycline and chloramphenicol).
(4) Preparation of Mobile Phase A solution
0.01 mol/L sodium dihydrogen phosphate plus 0.001 mol/L EDTA, adjusted to pH 2.0 with 30% nitric acid, and filtered through a 0.45 μm double-layer aqueous membrane filter.
(5) Preparation of Mobile Phase B solution
0.01 mol/L sodium dihydrogen phosphate, adjusted to pH 2.0 with 30% nitric acid, and filtered through a 0.45 μm double-layer aqueous membrane filter.
(6) Sample Pre-treatment
Accurately weigh 5.00 g of minced aquatic products into a 50 ml plastic centrifuge tube with a cap. Add 15 ml of 20% perchloric acid solution, vortex for 1 minute, and then place the centrifuge tube into an ultrasonic cleaner for extraction for 20 minutes. Subsequently, centrifuge at 4,000 rpm for 10 minutes, and collect the supernatant into a 25 ml volumetric flask. Add 10 ml of 2% perchloric acid solution to the residue, and use a glass rod to stir thoroughly until the sample is uniformly dispersed in the perchloric acid solution. Place this centrifuge tube into an ultrasonic cleaner for extraction for 20 minutes, then centrifuge at 4,000 rpm for 10 minutes, and collect the supernatant into the same 25 ml volumetric flask. Combine the two supernatants, make up to volume with 2% perchloric acid solution, and take 10 ml of this extract for adsorption onto a pre-conditioned solid-phase extraction column. Elute the solid-phase extraction column with 2 ml of 10% methanol-water, followed by 1 ml of 50% methanol-water, and collect the eluate for later use.
(7) Sample determination