In addition to balancing the centrifuge, you should also pay attention to sealing.
Release time:
2020-11-03
Overview:
Centrifuge Sealing is not merely a simple seal—it’s also driven by safety considerations.
During centrifugation, if the seal is not tight enough, at best it will lead to leakage of the separation medium; if the medium is corrosive, it could damage the equipment and cause the separation experiment to fail. At worst, it could result in a serious experimental accident that poses a threat to human life.
Moreover, in the medical and pharmaceutical industries, whether operating under non-sterile or sterile conditions, centrifuges place extremely high demands on the surrounding environment during use. Under higher-grade environmental requirements—for example, in sterile workshops—workshops are classified into levels such as Class 10,000 and Class 100,000. Excellent sealing performance ensures effective isolation between the external environment and the machine’s interior.
Hood seal
The hood should be hinged and equipped with reliable sealing measures at the joint with the housing. If the sealing structure or material fails to provide adequate corrosion resistance, it could lead to leaks of liquid or gaseous substances, posing risks to the environment and human health.
When designing and selecting sealing groove structures and sealing strips, their sealing performance and reliability should be fully considered.
Reliable sealing measures should also be provided for the installation interfaces of additional components such as feed pipes, washing pipes, and observation windows.
Bearing seat seal
Regarding the sealing of the spindle nut and bearing housing, earlier conventional models featured relatively simple sealing measures in these areas, achieving only partial sealing. For instance, at the spindle nut location—the mating surface between the spindle nut and the rotating drum—sealing rings are typically not installed. As a result, during operation, liquid and gas phases can seep into this area, causing corrosion on the tapered surface of the spindle and even penetrating into the bearings, thereby damaging them and compromising the normal operation of the machine. The upper bearing cover usually incorporates structures such as skeleton seals; however, a portion of the space between the upper bearing cover and the conical bore of the rotating drum remains exposed. Consequently, the gas phase within the centrifuge’s internal cavity may corrode the exposed surfaces of the spindle. Since centrifuges used in the pharmaceutical industry often handle materials containing organic solvents or media such as acids, bases, and salts, it is crucial to take full account of the sealing requirements for these critical areas when ordering equipment and to incorporate either dynamic or static sealing structures to meet the necessary sealing standards.
In addition to sealing the critical points, there’s also balancing—mentioned repeatedly in earlier articles. These two aspects are of paramount importance. Furthermore, during the operation of a centrifuge, you must maintain a keen and alert mind, constantly being vigilant for any unusual signs or anomalies.
For example, intense vibration and loud noise:
Check for any imbalance issues. Verify that the nuts securing the machine are not loose; if they are, tighten them securely. Inspect the bearings for damage or bending; if either is found, replace the bearings. If the machine’s outer cover is deformed or improperly positioned and causing friction, adjust it accordingly.