There are many factors that affect the performance of centrifuges.

Release time:

2017-06-23


Overview:

   Centrifuge It is a high-tech piece of equipment. As a solid-control device for drilling fluids, the centrifuge, when properly designed and used, can effectively control the density of drilling fluids and remove harmful solid phases from them. This plays a crucial role in enhancing mechanical drilling rates, extending the service life of downstream equipment, reducing the consumption of mud additives, and protecting oil and gas reservoirs. In China, medium-speed centrifuges began to be used only in the mid-1990s, and it was then recognized that centrifuges have tremendous potential in controlling solid phases in drilling fluids—especially in certain special-purpose wells where their effectiveness is even greater. As drilling operations become increasingly challenging and the demands on solid-control systems continue to rise, medium-speed centrifuges alone can no longer adequately address today’s challenges in drilling-fluid solid-phase control. To keep pace with the development of new drilling technologies and further improve the level of solid-control in drilling fluids, it is essential to enhance the performance of centrifuges—particularly by developing high-speed centrifuges with large throughput and high reliability. Many factors influence the performance of centrifuges, and the selection of structural parameters is one of the most critical aspects. Such selections must strike a careful balance among various factors, taking into account both the requirements of the centrifugal separation process and economic considerations.
  According to conventional theory, centrifuge malfunctions are caused by excessive rotor imbalance. The main factors contributing to rotor imbalance include weighing errors in samples, spillage of centrifuged sample liquids during the centrifugation process, and rotor corrosion.
  The main reasons are as follows:
  1. During centrifuge tube balancing, due to prolonged use of the balance without calibration, the symmetrically placed centrifuge tubes became unbalanced by more than 0.1 to 1 gram.
  2. When loading samples into centrifuge tubes, if the sample liquid fills the tube completely or the tube cap is not tightly screwed on to seal it properly, then... Centrifuge of the During centrifugation, the high-vacuum state within the centrifuge chamber can cause centrifuge tubes to flatten, rupture, and leak sample liquid, leading to rotor imbalance and potentially resulting in shaft bending or shaft breakage accidents.
  3. An accident occurred because the specific gravity of the original sample was not equal to that of the balancing liquid, causing rotor dynamic imbalance.
  4. When using a horizontal rotor, failing to carefully match the serial number of the bucket with the serial number on the rotor body can affect the dynamic balance of the centrifuge rotor.
  5. Accidents occur when aluminum alloy centrifuge tube caps and stainless steel centrifuge tube caps are used interchangeably (due to their different specific gravities).
  6. Improper use and inadequate sterilization of the rubber sealing rings inside the centrifuge tube caps and rotor cover—such as prolonged use leading to internal fractures, or exposure to high-temperature sterilization (e.g., baking in a drying oven)—can cause these rings to age, crack, and lose their sealing effectiveness. As a result, samples may spill out during high-speed operation, causing the rotor to run in an unbalanced state.
  7. For centrifuge tubes made of various materials, failure to thoroughly understand the manufacturer’s specified usage limits and disinfection methods before use led to improper handling with chemical solvents and inappropriate disinfectants, resulting in swelling and rupture of the centrifuge tubes during operation and subsequent accidents.
  8. Due to carelessness during work, the rotor cover was not tightened properly or the rotor cover was mistakenly swapped with the rotor shaft, resulting in mismatched screw threads. When the machine was started and operated, the rotor cover flew off, causing a severe shaft breakage accident.
  The centrifuge industry has tremendous growth potential, and we need to seize this opportunity effectively. However, when choosing which one to use... Centrifuge At this time, be sure to follow the basic instructions. After turning on the power, the drum will rotate clockwise. Place 8–10 layers of filter paper into the centrifuge chamber, then lay a filter cloth on top. Rinse the filter cloth and the machine with water, repeating this process three times before proceeding to centrifuge the drug.