Software Operation and Debugging of Laboratory High-Speed Refrigerated Centrifuges

Release time:

2012-05-09


Overview:

A high-speed refrigerated centrifuge is primarily used to extract the desired components from liquid mixtures. By virtue of the different densities of each substance, as the mixture undergoes high-speed rotation, the denser liquids settle at the bottom, while the less dense liquids float on top, thereby separating the liquid into distinct layers and allowing us to isolate the pure substances we need.

A high-speed refrigerated centrifuge is primarily used to extract the desired components from liquid mixtures. By means of high-speed rotation, substances with different densities separate: the denser liquids settle at the bottom, while the less dense liquids float on top, thus layering the liquid and allowing us to isolate the pure substances we need. It finds applications in many fields, including pharmaceuticals and chemical engineering.


1. High-speed refrigerated centrifuge software procedure

During prototype debugging, control subroutines for seven rotor models have been identified, suitable for rotors operating at speeds ranging from several thousand to thirty thousand revolutions per minute. In actual application, users only need to enter the rotor model they are using via the keyboard.

2 System Operation Results

In the event of control failures or mechanical system malfunctions, the system should be able to detect them promptly and trigger an alarm, shutting down all controlled devices to minimize system damage. At the same time, when detecting faults that do not affect production—such as damaged buttons or abnormal displays—the system should issue an alarm to alert the operator without halting production.

3. Issues Encountered During Debugging and Their Solutions

(1) Since there is a power-frequency quenching furnace and an electric welding workshop adjacent to this production workshop, industrial interference is relatively severe, significantly impacting the system’s normal operation and frequently causing the program to “run out of control.” To address this issue, we have implemented several anti-interference measures, including:

a. Software anti-interference measures;

b. In addition to implementing white reset measures, the hardware has also been further enhanced with shielding and isolation measures.

c. Use an independent power supply (with phase separation from the interference source) or install a UPS power supply;

d. An IRC absorption circuit was added to the relay coil. With these measures in place, the system has been functioning normally ever since and has not experienced any further crashes.

(2) The sampled signal was unstable, which affected the reliability of the control. Due to severe interference, the stability of the upstream signals was compromised, leading to spurious phenomena in the sampled signal and causing false operations. Based on our analysis of the phenomena observed on-site, we ultimately resolved this issue by adding a hardware filtering circuit to the preceding stage.