Maintenance of Speed Faults in Desktop Centrifuges
Release time:
2012-01-12
Overview:
During the experiment with a desktop centrifuge, the motor speed may be too slow and the carbon brushes may produce significant sparking. In such a situation, what steps should be taken?
Our centrifuge manufacturer’s technical engineers will answer your questions one by one.
I. Analysis and Troubleshooting: This fault is caused by dirt or reduced spring tension in the carbon brush, resulting in poor contact between the carbon brush and the commutator segments.
II. Solution: Disassemble the machine, unscrew the brush holder cover, and remove the carbon brushes along with their springs. First, clean the spring, carbon brushes, and commutator segments with anhydrous alcohol to remove any dirt or debris. If there are chipped or damaged areas at the front end of the carbon brushes, use a combination file together with sandpaper to carefully grind these areas into smooth, curved contact surfaces that fit snugly against the round commutator segments. Then, appropriately increase the spring tension.
After the above-mentioned troubleshooting steps, the fault can usually be resolved. For after-sales service support, please call 021-68113426.
This article is provided by Shanghai Luxiangyi Centrifuge Instrument Co., Ltd. Please cite the source when reprinting. http://www.lxylxj.com