Addressing the issue of water accumulation inside the chamber of low-speed, high-capacity centrifuges.
Release time:
2012-01-12
Overview:
When a refrigerated centrifuge is running, the refrigeration causes moisture in the air to condense into frost inside the centrifuge chamber. After the machine is turned off, this frost melts into water. Most low-speed, high-capacity centrifuges from overseas do not have drain holes, so water gradually accumulates in the centrifuge chamber. In such cases, users should remove the rotor themselves and clear out the accumulated water. When reinstalling the rotor, be sure to secure it properly to prevent accidents.
Aluminum alloys are not susceptible to corrosion.
Centrifuge rotors for freezing are typically made of aluminum alloy. Once corroded, their strength diminishes, making them prone to accidents. Aluminum alloys are susceptible to liquid corrosion; after cleaning, they should be dried thoroughly using a hairdryer or by turning them upside down for a period of time—only use them once you’ve confirmed they’re completely dry. Some blood banks import medical centrifuges that lack the six plastic holders that should normally be included in the six aluminum cups. These plastic holders are meant to hold blood bags, preventing direct contact between the bags and the aluminum cups. Also, be sure to check test tubes carefully for any cracks.
Handling of Shaft Breakage Accidents
If the rotor is not properly installed or if the sample loading is severely unbalanced, the centrifuge will be started. Although modern imported centrifuges are equipped with imbalance protection—meaning they will automatically cut power and stop the rotor once the imbalance exceeds a certain threshold—it’s already too late by then, and a serious accident could occur. One such severe accident is shaft failure: when the shaft suddenly breaks during high-speed rotation, the rotor loses its support and starts spinning wildly, colliding randomly within the centrifuge chamber. This can cause the entire centrifuge unit to rotate through an angle of 270 degrees.
The above information was compiled and edited by Shanghai Luxiangyi Centrifuge Instrument Co., Ltd. Editor: admin