Technique for separating red blood cells using a low-speed centrifuge

Release time:

2011-12-28


Overview:

Centrifuge the sample using a horizontal rotor centrifuge at 2,000 rpm for 20 minutes. After centrifugation, you will observe that the sample has separated into multiple layers: the bottom layer consists of red blood cells, the middle layer is the layered liquid, and the top layer is plasma. Between the plasma layer and the layered liquid, there is a thin, relatively dense white layer—the mononuclear cell layer.

Materials and Reagents

1. Anticoagulants such as heparin 2. 3% Gelatin Solution: Take 3 g of gelatin and dissolve it in 100 ml of 0.9% sterile saline. Sterilize at 114.3°C for 10 minutes, then cool and store at 4°C in a refrigerator. Before use, preheat at 37°C for 10 minutes. 3. Alsever’s Solution

(2) Operating Procedures

1. Blood collection with anticoagulation yields red blood cells. Since the ratio of red blood cells to white blood cells is vastly different, white blood cells are typically mixed in at a negligible level and can be disregarded.
 
2. Depending on the intended use of the red blood cells, further processing may be required. For example, if you need to count red blood cells, you can directly use a diluted counting chamber. However, if you plan to perform a complement-binding assay or other red blood cell lysis reactions, the red blood cells must be thoroughly washed to remove plasma adhering to the surface of the red blood cell membrane. Typically, this is done by repeatedly washing the cells with an isotonic diluent, followed by centrifugation at 2,000 r/min for 10 minutes, and repeating this process 3 to 4 times.

3. If storage for future use is required, collect blood using ACD solution as the anticoagulant. After thorough mixing, store the sample at 4°C; it can be preserved for up to 2 weeks while still maintaining its activity.

Lymphocyte isolation techniques

(1) Materials and Reagents

1. There are two types of lymphocyte separation media:

(1) Polysucrose-Iodixanol Solution: Polysucrose solution, commercially known as Ficoll, has a molecular weight of 400,000. It is commonly prepared as an aqueous solution at a concentration of 40 ± 1% (w/v); it is also available in dry powder form. When using, prepare a 9% solution with distilled water. Iodixanol solution (Meglumini diatrizoici), commercially known as Vrografin, has the structural formula 3,5-diacetamido-2,4,6-triiodobenzoic acid 1-deoxy-methylamino-sorbitol. Its concentration is either 60% or 75%, and each vial contains 20 ml. It is frequently used for organ imaging in humans. To prepare the solution, take 20 ml of 60% iodixanol stock solution and add 15.38 ml of double-distilled water, resulting in a 33.9% iodixanol solution. Preparation of 1.077 ± 0.001 Polysucrose-Iodixanol Solution: 9% Polysucrose Solution: 24 parts 33.9% Iodixanol Solution: 10 parts Mix thoroughly. If necessary, measure the specific gravity. For storage, filter through a G5 funnel to remove bacteria or sterilize by autoclaving at 114.3°C for 15 minutes, then store at 4°C. Typically, this solution can be stored for up to 3 months. If you need to prepare layered solutions with different proportions, use the following formula: In this formula, dm represents the specific gravity of the layered solution, d1 and d2 represent the specific gravities of the 9% polysucrose and 33.9% iodixanol solutions, respectively, and V1 and V2 represent their respective volumes. For example, if you want to prepare a layered solution with a specific gravity of 1.077 using polysucrose and iodixanol, the ratio of 9% polysucrose to 33.9% iodixanol should be 100:46.3.

(2) Iopamidol-Dextran Solution: Mix 10 parts of 34% iopamidol solution with 12.5 parts of 60% dextran solution to obtain a stratified liquid with a specific gravity of 1.07–1.09. Dispense into brown bottles and store at 4°C for future use. 2. 3% Gelatin Solution: Weigh 3 g of gelatin, dissolve it in 100 ml of sterile 0.9% saline solution, and autoclave at 114.3°C for 10 minutes. After cooling, store the solution at 4°C; before use, preheat it at 37°C for 10 minutes. 3. Hank’s Solution.

(2) Operating Procedures

1. Collect anticoagulated blood and allow it to settle naturally. If the blood is from equine species, you can directly place the tubes upright in a tube rack and let them settle naturally for 1 hour; then carefully collect the upper layer of plasma. If the blood is from cattle or sheep, since their erythrocyte sedimentation rate is very slow, you can add an equal volume of 3% gelatin solution, mix thoroughly, and allow the mixture to settle naturally. After 1 hour, collect the upper layer of plasma.
 
2. Centrifuge at 2,000 rpm for 10 min, then discard the supernatant.
 
3. After suspending the precipitate in Hank’s solution, centrifuge at 2,000 r/min for 10 minutes.
 
4. Repeat Step 3 once more, then use the cell nutrient solution to prepare a suspension of white blood cells. This solution contains the entire population of white blood cells, including granulocytes, lymphocytes, monocytes, and a portion of red blood cells. It is suitable for white blood cell counting.

5. Transfer 2 ml of cell separation medium into a centrifuge tube. At the same time, use a capillary pipette to carefully add approximately 2 ml of plasma (the upper layer of plasma that has been allowed to settle naturally for 1 hour) onto the surface of the separation medium. Alternatively, you can directly add anticoagulated blood.
 
6. Centrifuge using a horizontal rotor centrifuge at 2,000 rpm for 20 minutes. After centrifugation, you will observe that the sample has separated into multiple layers: the bottom layer consists of red blood cells, the middle layer is the interphase liquid, and the top layer is plasma. Between the plasma layer and the interphase liquid, there is a thin, relatively dense white layer—the mononuclear cell layer.
 
7. Insert a capillary pipette into the mononuclear cell layer and aspirate this layer, then transfer it into another test tube.

8. Wash and centrifuge using Hank’s solution, then adjust the suspension to an appropriate white blood cell concentration. Count the cells if necessary.

(3) Precautions

1. When adding plasma or blood to the separation medium, proceed with caution and slowly, avoiding disruption of the liquid layers and refraining from shaking. Alternatively, you can add the separation medium to the upper layer of the plasma.
 
2. Maintaining the viability of lymphocytes is extremely important; therefore, in most cases, blood is collected fresh and immediately processed for separation.
 
3. The lymphocyte layer obtained after separation by density gradient centrifugation is actually a mononuclear cell layer, which includes monocytes but excludes granulocytes. If you wish to isolate pure lymphocytes, you can remove the monocytes according to the following procedure, or harvest the monocytes directly.