Common precautions for centrifuges in different laboratories
Release time:
2020-07-24
Overview:
Laboratory centrifuge It’s a broad term—why is that? Because “laboratory centrifuge” is a generic term that can refer to high-speed centrifuges, low-speed centrifuges, refrigerated centrifuges, and many other types; all of these can be classified as laboratory centrifuges.
Centrifugation is a fundamental laboratory procedure that uses a centrifuge to separate components from complex mixtures.
When centrifuging a sample at extremely high speeds, the components of the mixture are subjected to centrifugal force, causing particles with higher densities to move away from the axis of rotation, while particles with lower densities move toward the axis.
These particles will settle at the bottom of the tube, forming what is known as sediment. The sample separated in this way—or the supernatant liquid remaining—can be used for further processing or analysis.
However, here’s an important point: Before using the centrifuge, you’ll need to convert RCF and RPM yourself using the appropriate formula.
The main component of a centrifuge is the rotor, which is the rotating part that spins at high speed during centrifugation.
The rotor can be fixed to the centrifuge, or a single centrifuge can use multiple rotors, which are mounted onto a component known as the spindle.
Centrifuge rotors usually come with a lid that can be tightly screwed on to prevent the sample from flying out.
Many centrifuges are equipped with a refrigeration unit that can maintain controlled internal temperatures during centrifugation.
The centrifuge also has knobs or buttons for entering the parameters required for each spin:
This includes centrifugation time, temperature, and centrifugation speed, which can be expressed either in terms of speed or relative centrifugal force.
Centrifugal speed is measured in RPM, which stands for revolutions per minute. The RPM value can sometimes be misleading, because what actually causes particles to separate from the mixture is not the speed of the centrifuge itself, but rather the force acting on those particles.
This force is related to the radius of the centrifuge rotor. Since different centrifuges have rotors of varying sizes, the force generated at the same RPM will differ accordingly.
Centrifugal force can also be quantified using relative centrifugal force (RCF).
The RCF typically manifests as a multiple of Earth's gravitational acceleration.
The formula for calculating the RCF is the rotor radius multiplied by the square of the angular velocity, then divided by Earth's gravitational acceleration.
The RCF is related to RPM via the following formula, where r represents the radius of the centrifuge, in centimeters.
The use of this formula may affect the success or failure of your experiment, but you don’t need to adopt this calculation method for every new operation. Centrifuges usually come with charts that make it easy to convert between RCF and RPM. Simply use a ruler to connect the centrifuge’s radius to a known RPM value, and you’ll obtain the corresponding RCF value.
Converting between RCF and RPM is essential for every centrifuge. Although most centrifuge manufacturers now provide charts to help easily convert between RCF and RPM, centrifuge operators still need to be able to perform the calculations themselves for verification purposes, thus avoiding errors.