How should vibration be controlled when using a high-speed centrifuge?
Release time:
2013-01-15
Overview:
The low-speed centrifuge is one of the most commonly used laboratory instruments. So, how should we control its vibration range during operation? As we all know, the rotor can only rotate smoothly if the centrifuge’s operating speed is not at or near its critical speed. For low-speed centrifuges with relatively low operating speeds, the critical speed is generally raised as much as possible, ensuring that the centrifuge operates below its critical speed. Such centrifuges have thick shafts with high rigidity, and these shafts are referred to as rigid shafts. When higher rotational speeds are required, the critical speed is minimized as much as possible, and the shaft is made thinner with lower rigidity—these are called flexible shafts. Modern ultracentrifuges all employ flexible shafts.
The critical speed is related to both the centrifuge itself and the rotor it is equipped with. A single centrifuge can be fitted with several different types of rotors; therefore, when designing a centrifuge, it’s essential to take into account the vibration-damping design’s compatibility with various rotors. Centrifuges operate over a wide range of speeds—from several thousand to tens of thousands of revolutions per minute. The higher the rotational speed at which vibration occurs in a centrifuge, the greater the associated risk.
Once vibration occurs at high rotational speed, the centrifuge should be immediately brought to a stop using the fastest deceleration gear. All personnel must promptly evacuate the area to avoid injury in case of accidents such as rotor breakage. Under no circumstances should the power be cut off—once the power is interrupted, the braking mechanism (which provides rapid deceleration) will fail, and the rotor will come to a halt very slowly. Such accidents are typically caused by improper operation of the centrifuge—for example, if the rotor is not properly seated on the spindle, the rotor lid is not securely closed, or the basket for swinging out the rotor is not positioned correctly. These accidents are accompanied by a loud bang; a Shanghai centrifuge weighing 300–400 kg may suddenly spin over 270 degrees. Upon opening the lid, you’ll find the rotor lying inside the chamber. When the rotor is removed, you’ll notice that its outer surface has been severely scratched and discolored. The inner walls of the centrifuge chamber will also be damaged, and in some cases, refrigerant may even leak out. Occasionally, the rotor and its lid separate, causing liquid from the tubes to spill into the chamber; in extreme cases, the spindle may break or bend severely.
Accidents caused by improper user operation occur repeatedly. The Lu Xiangyi centrifuge from Shanghai is equipped with an imbalance protection feature: once the rotor's vibration exceeds a certain threshold upon startup, the centrifuge automatically slows down and shuts down. However, centrifuges with this feature sometimes fail to trigger the alarm correctly, and thus cannot completely guarantee that accidents will be avoided.