Centrifuge In fact, there are many details in the operation that you need to understand—not just the issue of balancing equations. Of course, balancing equations is both the foundation and the key focus.
The laboratory and production environments constantly remind us that, at present, accidents involving centrifuge experiments due to improper operation still occur from time to time. Therefore, it is essential to cultivate awareness of safe centrifuge operation. The following provides further details on the fundamentals of safe centrifuge operation.
(1) Precisely balance the centrifuge tube and its contents.
1. Balancing Centrifuge Tubes and Their Contents: When using various centrifuges, it is essential to precisely balance all centrifuge tubes, tube holders, caps and top covers, shields, and tube sleeves on a balance beforehand. The weight difference during balancing must not exceed the range specified in the instruction manual for each particular centrifuge. Tube sleeves and shields are usually labeled with their weights; be careful not to mix up matching tube sleeves and shields during the balancing process. The tubes must be filled with materials that are similar to those intended for centrifugation—for example, when centrifuging bacteria from a culture medium, you can use the medium itself for balancing. However, you must never use water to balance tubes containing cesium chloride.
2. Symmetrical loading of centrifuge tubes: An odd number of centrifuge tubes is not allowed in the rotor. When the rotor is only partially loaded, place the tubes symmetrically within the rotor to ensure that the load is evenly distributed around its circumference.
(2) Properly Load the Solution: When loading the solution, follow the specific operating instructions for each type of centrifuge. Select appropriate centrifuge tubes based on the properties and volume of the solution to be centrifuged. Some centrifuge tubes are open-top; do not overfill them with liquid, as this could cause the liquid to splash out during centrifugation, leading to rotor imbalance, rusting, or corrosion. For ultracentrifuge tubes used in preparative ultracentrifuges, it is often required to fill the tubes completely to prevent the upper part of the plastic centrifuge tubes from collapsing and deforming during centrifugation.
(3) Ensure the lid is securely closed before starting: Before initiating the centrifuge, always remember to close the laboratory benchtop centrifuge chamber door or lid as well as the rotor lid tightly. After starting the centrifuge, the operator must not leave the machine until the rotation speed has reached the preset value and the operation has stabilized. However, the operator should continue to monitor the running condition at all times. During operation, if any abnormal situation occurs, immediately stop the centrifuge and take appropriate corrective measures. In the event of a sudden power failure during operation, you must first disconnect the power supply and wait for the rotor to gradually slow down due to inertia. Once it has come to a complete stop, manually open the centrifuge chamber door and remove the samples and rotor.
(4) Remove the centrifuge tubes promptly: Immediately after the centrifugation is complete, remove the centrifuge tubes from the centrifuge. Once the contents of the tubes have been transferred out, promptly clean the tubes and allow them to air-dry naturally.
(5) Clean the rotor promptly: After centrifugation is complete, the rotor should be washed with warm water and dried. Generally, rinsing with water alone is sufficient; never soak the rotor in detergents. Once cleaned, dry the rotor thoroughly and allow it to air-dry in a clean, dry environment. If the rotor will not be used for an extended period, apply a layer of polishing wax for protection.
Now that we’ve finished discussing centrifuge balancing, let’s move on to cover other key points, such as the rotor, rotational speed, separation time, and temperature.
Rotor service life:
For domestically produced high-speed and low-speed centrifuge rotors, rotor lifespan is simply specified based on years of service; aluminum alloy rotors typically have a lifespan of 7 years. This approach is clearly not very scientific. Some manufacturers, therefore, also keep track of the number of operating cycles or total operating hours. If a particular rotor is used almost every day for 20 hours at a time, it certainly won't last the full 7 years. On the other hand, if a centrifuge has been purchased many years ago but hasn't been used at all, yet has reached the manufacturer’s prescribed 7-year mark, our factory technicians generally find that such rotors are still usable—though they may need to operate at a slightly reduced speed.
Centrifuge speed: The magnitude of centrifugal force depends on the rotor’s rotational speed and the centrifugal radius. When specifying centrifugation conditions, low-speed centrifugation is typically expressed in terms of revolutions per minute (rpm) of the rotor. In high-speed centrifugation—especially ultrahigh-speed centrifugation—relative centrifugal force (RCF) is often used instead. In practical applications, the centrifugal force usually refers to the average value of the relative centrifugal force, that is, the centrifugal force at the midpoint of the centrifuge tube.
Separation time: The separation time varies depending on the centrifugation method used. For differential centrifugation, the separation time refers to the duration required for particles to completely settle to the bottom of the centrifuge tube.
For rate-zonal centrifugation, the separation time refers to the duration required to form clearly defined zones.
For isopycnic density gradient centrifugation, the separation time refers to the equilibrium time required for particles to completely reach the isopycnic point.
The factors influencing the band formation time required for rate-zonal centrifugation and the equilibration time required for isopycnic band centrifugation are complex and can be determined through experimentation.
Separation temperature: To prevent the aggregation, denaturation, and inactivation of separated substances, in addition to carefully selecting the centrifugation medium, it is also essential to maintain an appropriate separation temperature—typically around 4°C. For substances with good thermal stability, centrifugation can be carried out at room temperature. During high-speed or ultrahigh-speed centrifugation, the rapid rotation of the rotor generates heat, leading to a rise in temperature; therefore, a cooling system must be employed to keep the temperature within a specified range.
Finally, let me emphasize that attention should be paid to corrosion issues in the localized components of centrifuges. Localized corrosion is prevalent both at high and low speeds, and it’s crucial to analyze its formation mechanisms carefully and take appropriate measures for maintenance and protection.
1. Pitting corrosion is a highly corrosive phenomenon that primarily occurs in metals prone to passivation, such as stainless steel. It arises because of potential defects localized on the surface, combined with the presence of active ions in the solution that can disrupt the passive film. As the passive film is locally compromised, an electrochemical couple is formed, leading to pitting corrosion. Once pitting corrosion has initiated, the centrifugal force accelerates the dynamic processes within the pit, ensuring the continued progression of corrosion until complete perforation occurs—this behavior differs from pitting corrosion under static conditions.
2. Wear-corrosion refers to the simultaneous damage caused by both wear and corrosion on a part’s surface.
3. Intergranular corrosion can cause parts to lose strength and ductility, leading to brittle fracture. It is a hazardous form of damage that develops inward from the surface along grain boundaries, with no visible signs of corrosion on the outer surface.
4. Stress corrosion is damage caused by the action of a certain tensile stress in a corrosive environment. It exhibits the following characteristics: residual tensile stress, applied external tensile stress, a corrosive and penetrating environment, and localized defects.
In fact, as long as proper daily maintenance is carried out, these issues are unlikely to arise. However, unforeseen circumstances can always occur. If they do, please don't disassemble the centrifuge yourself—instead, promptly contact the centrifuge manufacturer and seek their advice and guidance through frequent communication.