Application of Whole-Blood Genomic DNA Extraction Using a Centrifuge
Release time:
2014-10-13
Overview:
Centrifuge-based genomic DNA extraction from whole blood
Operating Procedure (200 µl whole blood)
# Before first use, please add 60 ml of anhydrous ethanol to 15 ml of wash solution WB!
1. Take 200 µl of fresh, frozen, or blood samples preserved with various anticoagulants and place them into a 1.5 ml centrifuge tube.
If the initial whole blood volume is less than 200 μl, use Buffer BB to bring the volume up to 200 μl. If the initial volume is between 200 μl and 300 μl, the subsequent steps require proportionally increasing the reagent volumes. If the initial volume is between 300 μl and 1 ml, you must first perform red blood cell lysis (see the appendix at the end of this manual).
2. Add 200 μl of binding solution CB, immediately invert and gently shake vigorously to ensure thorough mixing. Then add 20 μl of proteinase K solution (20 mg/ml), invert and gently shake again to ensure thorough mixing. Incubate at 70℃ for 10 minutes; the solution should become clear and bright (though its color may tend toward black).
3. Add 100 μl of isopropanol, invert vigorously and gently shake to mix thoroughly. At this point, a flocculent precipitate may appear.
In the above-mentioned operational steps, it is crucial to thoroughly mix with appropriate force. Inadequate mixing can significantly reduce yield. If the sample is viscous and difficult to mix, you may vortex it for 15 seconds to achieve uniform mixing; however, do not vigorously shake it by hand, as this could shear the DNA.
4. Add both the solution obtained in the previous step and the flocculent precipitate into an adsorption column AC (place the adsorption column into a collection tube). Centrifuge at 10,000 rpm for 30 seconds, then discard the waste liquid from the collection tube.
5. Add 500 μl of Inhibitor Removal Solution IR, centrifuge at 12,000 rpm for 30 seconds, and discard the supernatant.
6. Add 700 μl of wash buffer WB (please first check whether anhydrous ethanol has been added!). Centrifuge at 12,000 rpm for 30 seconds, then discard the waste liquid.
7. Add 500 μl of wash buffer WB, centrifuge at 12,000 rpm for 30 seconds, and discard the supernatant.
8. Place the adsorption column AC back into the empty collection tube, and centrifuge at 13,000 rpm for 2 minutes to remove as much of the wash solution as possible, thereby preventing residual ethanol in the wash solution from inhibiting downstream reactions.
9. Remove the adsorption column AC and place it into a clean centrifuge tube. Add 100 μl of elution buffer EB—preheated to 65–70℃—in the middle of the adsorption membrane. Allow it to stand at room temperature for 3–5 minutes, then centrifuge at 12,000 rpm for 1 minute. Transfer the resulting solution back into the centrifugal adsorption column, let it stand at room temperature for 2 minutes, and then centrifuge again at 12,000 rpm for 1 minute. The larger the elution volume, the higher the elution efficiency. If a higher DNA concentration is required, you can appropriately reduce the elution volume; however, the minimum volume should not be less than 50 μl. A volume that is too small will decrease the DNA elution efficiency and reduce the DNA yield.
10. DNA can be stored at 2–8℃; for long-term storage, it can be kept at -20℃.
Operating Procedure (300 µl and 1 ml whole blood)
1. Pipette 900 μl of red blood cell lysis buffer (available for purchase from our company) into a 1.5 ml centrifuge tube, or 3 ml of red blood cell lysis buffer into a 15 ml centrifuge tube.
2. After gently inverting and mixing the anticoagulated whole blood (which should be brought back to room temperature before use), pipette 300 μl of whole blood and 1 ml of whole blood into the 1.5 ml and 15 ml centrifuge tubes mentioned above, respectively. Invert the tubes 6–8 times and then gently tap the tube walls upside down to ensure thorough mixing.
3. Allow to stand at room temperature for 10 minutes (during this time, gently invert and tap several times to help lyse the red blood cells).
4. Centrifuge at 12,000 rpm for 20 seconds (for 1.5 ml centrifuge tubes), or at 2,000–3,000 rpm for 5 minutes (for 15 ml centrifuge tubes). Discard the red supernatant, and carefully aspirate as much of the supernatant as possible without disturbing the cell pellet at the bottom of the tube. Leave behind a complete white blood cell pellet at the bottom of the tube, along with approximately 10 μl of residual supernatant. After centrifugation, you should observe a white white blood cell pellet at the bottom of the tube; there may also be some red blood cell fragments mixed with the white blood cell pellet. However, if most of the pellet appears red, it indicates that the red blood cell lysis was insufficient. In this case, add more red blood cell lysis buffer to resuspend the cell pellet and repeat steps 3 and 4.
5. Add 200 μl of Buffer BB, vortex and gently shake to resuspend the white blood cell aggregates, ensuring thorough dispersion of the aggregates. Note that white blood cell aggregates precipitated from heparin-anticoagulated blood are particularly difficult to disaggregate and resuspend, which may compromise the lysis efficiency in subsequent experiments. Therefore, we recommend using an anticoagulant other than heparin when collecting blood samples.
6. You can now extract DNA according to the operational steps.
Operating Procedure (10 ml whole blood)
1. Pipette 30 ml of red blood cell lysis solution into a 50 ml centrifuge tube.
2. After inverting and mixing the anticoagulated whole blood (which should be brought back to room temperature before use), pipette 10 ml into the centrifuge tube containing the red cell lysis solution mentioned above. Invert the tube 6–8 times and gently tap the tube wall while holding it upside down to ensure thorough mixing.
3. Allow to stand at room temperature for 10 minutes (during this time, gently invert and tap several times or gently shake on a shaker to help lyse the red blood cells).
4. Centrifuge at 2,000 × g for 10 minutes. Discard the red supernatant, and carefully aspirate as much of the supernatant as possible (taking care not to disturb the cell pellet at the bottom of the tube). Leave behind a complete white blood cell pellet at the bottom of the tube along with approximately 100 μl of residual supernatant. After centrifugation, you should observe a white-white blood cell pellet at the bottom of the tube; there may also be some red blood cell fragments mixed with the white blood cell pellet. However, if most of the material at the bottom is red, it indicates that the red blood cell lysis was insufficient. In this case, add an appropriate amount of red blood cell lysis buffer to resuspend the cell pellet and then repeat steps 3 and 4.
5. Vortex the suspension for 15 seconds to resuspend and thoroughly disperse the white blood cell aggregates. The resuspension and dispersion of white blood cells are crucial for the subsequent lysis step. If the white blood cells are not fully dispersed before adding the lysis buffer, they may fail to lyse adequately, resulting in visible clumps that can be observed with the naked eye.
6. Add 10 ml of nuclear lysis buffer to the resuspended leukocytes, and quickly and vigorously pipette up and down several times to ensure thorough mixing and lysis of the leukocytes. As genomic DNA is immediately released, the mixture will rapidly become highly viscous. At this point, immediately stop pipetting (to avoid shearing the genomic DNA). Invert and gently rotate the centrifuge tube 10 times to ensure that the lysis buffer comes into contact with all the leukocytes and achieves complete lysis. This step is critical for obtaining satisfactory lysis efficiency and maintaining genomic DNA integrity. If the pipetting force is too weak, it may not be sufficient to quickly disaggregate the cell clumps and thoroughly mix them with the lysis buffer, resulting in visible clumps that cannot be completely lysed. (If the lysis buffer contacts only the cell clumps rather than individual dispersed cells, it will immediately form a viscous barrier on the surface of the clumps, making it difficult for the buffer to penetrate further into the interior.) Once the lysis buffer comes into contact with the cells, genomic DNA is immediately released. At this stage, excessive pipetting force can easily lead to genomic DNA fragmentation. The correct approach is to quickly and vigorously pipette up and down several times; after just a few seconds, when the mixture becomes viscous due to the release of genomic DNA, immediately stop pipetting or switch to using a larger-caliber pipette tip (cutting off the tip’s sharp end) and gently pipette or invert the tube to mix. If visible clumps still remain, incubate the sample at 65°C for 30–60 minutes (but no longer than one hour) until complete lysis is achieved.
7. Optional step: Add RNase A (10 mg/ml) to the lysate to a final concentration of 30 μg/ml, invert the mixture 25 times to ensure thorough mixing, and incubate at 37℃ for 15 minutes to remove any residual RNA. Then, cool the mixture back to room temperature.
8. After adding 3.33 ml of protein precipitation solution, vortex the mixture vigorously and continuously on a vortex mixer for 20–25 seconds. After mixing, you may observe some small protein aggregates. Given the small volume and weight of the sample, the shear forces generated by vortexing will not shear or break down genomic DNA. If you mix manually, avoid vigorous up-and-down shaking; instead, use gentle but consistent shaking to ensure thorough mixing. Otherwise, the genomic DNA could be sheared and broken. However, the shaking intensity must not be too low either—ensure that the viscous lysate is fully dispersed. Otherwise, the DNA won't separate effectively from the protein precipitate, and during centrifugation, it might co-precipitate with the proteins, leading to DNA loss or reduced yield. Moreover, insufficient mixing can also result in incomplete protein precipitation, leaving the final product heavily contaminated with proteins. Therefore, we recommend that beginners use a vortex mixer for mixing.
9. Centrifuge at 2,000 × g (the centrifugal force can be adjusted upward depending on the centrifugation effect—e.g., if the pellet is not tightly packed). Centrifuge for 10 minutes. At this point, you should be able to observe a dark-brown protein precipitate at the bottom of the tube; you may also see some protein precipitate floating on the surface of the liquid.
10. Carefully pipette the supernatant (approximately 10 ml) into a new 15-ml centrifuge tube. When pipetting the supernatant, be careful not to draw up any protein precipitates that have settled at the bottom of the tube or are floating on the surface of the liquid. If you accidentally transfer protein precipitates into the new centrifuge tube, you can centrifuge again for 2 minutes and then carefully collect the supernatant.
11. Add an equal volume of room-temperature isopropyl alcohol (about 10 ml), gently invert the mixture 30 times to mix thoroughly, or until a cotton-like (filamentous) white DNA precipitate appears.
Note that sometimes, when the cotton-like (filamentous) DNA precipitate is inverted and mixed, it may adhere to the lid or the tube opening and fail to come down even after inversion. Alternatively, bubbles may become trapped, causing the precipitate to float on the liquid surface. As a result, the operator might not be able to see the precipitate and mistakenly conclude that no DNA has been obtained. To resolve this issue, skip step 12 and proceed directly to centrifugation at 2,000 x g for 1 minute, discard the supernatant, and then continue with step 14.
12. Place the centrifuge tube vertically, allowing the white DNA precipitate to settle naturally at the bottom of the tube. Then, carefully aspirate as much of the supernatant as possible, taking care not to disturb the precipitate.
13. After adding 10 ml of 70% ethanol, invert the tube to mix thoroughly, then centrifuge at 2,000 × g for 1–3 minutes. A white DNA precipitate will be visible at the bottom of the tube; discard the supernatant.
14. Add 3 ml of 70% ethanol, invert the tube several times to wash the DNA precipitate, then centrifuge at 2,000 × g for 1 minute. Discard the supernatant (the precipitate is very loose—be careful not to discard the DNA precipitate). Invert the tube and gently tap it on absorbent paper to remove any residual ethanol. You can also carefully use a pipette tip to remove any residual ethanol around the pellet at the bottom of the tube and along the tube walls. Allow the precipitate to air-dry for a few minutes. Be careful not to overdry the precipitate, as this could make the DNA extremely difficult to dissolve. Also, avoid leaving too much ethanol behind, as residual ethanol might inhibit downstream reactions such as restriction enzyme digestion.
15. Add 600 μl of DNA dissolution solution (or adjust the volume according to the DNA concentration) to rehydrate and dissolve the DNA precipitate. Gently tap the tube wall to mix thoroughly. The sample can be incubated at 65℃ for 30–60 minutes (but not longer than one hour). Alternatively, the sample can be left overnight at room temperature or 4℃ for rehydration, gently tapping the tube wall from time to time to facilitate DNA rehydration.
16. DNA can be stored at 2–8℃; for long-term storage, it can be kept at -20℃.