Operating a centrifuge isn't as simple as just holding the instruction manual.
Release time:
2020-07-10
Overview:
Centrifuge In addition to following the instructions carefully, you should also predict the success or failure of the experiment by observing the operational procedures and physical indicators.
The separation time of a centrifuge varies depending on the centrifugation method used.
For differential centrifugation, the separation time refers to the duration required for particles to completely settle to the bottom of the centrifuge tube.
For rate-zonal centrifugation, the separation time refers to the duration required to form distinct bands.
For isopycnic density gradient centrifugation, the separation time refers to the equilibrium time required for particles to completely reach the isopycnic point.
The factors influencing the band formation time required for rate-zonal centrifugation and the equilibration time required for isopycnic band centrifugation are complex and can be determined through experimentation.
For a routine biochemical test, you’ll need to separate the serum. This is a low-speed centrifuge, operating at a speed of around 3,000 to 5,000 rpm. Typically, the process takes no more than 3 minutes.
Serum separation typically involves centrifugation on the day of blood collection; if stored appropriately, it can be kept for up to one day. Hemolysis usually occurs when blood is transferred from the syringe into the centrifuge tube too quickly and under excessive pressure after collection, or when the centrifugation speed is too high—hemolysis tends to occur at speeds above 8,000 rpm.
1. For the centrifugal separation of serum from specimens, the relative centrifugal force (RCF) of the centrifuge should be set at 1000g to 1200g, and the centrifugation time should be 5 to 10 minutes. Increasing either the relative centrifugal force or the centrifugation time can easily lead to hemolysis.
II. Handling of Hemolyzed Specimens: Hemolysis can lead to the release of various components from RBCs, thereby interfering with assay methods. Hemoglobin (Hb) levels ≥0.2 g/L indicate visible hemolysis.
1. Indirect Interference: The release of hemoglobin can cause indirect interference, as hemoglobin exhibits light absorption at wavelengths of 431 nm and 555 nm. If the analyte being measured is analyzed colorimetrically using wavelengths close to these, it may lead to falsely elevated results. Treatment methods: For mild hemolysis, perform a serum blank simultaneously; for severe hemolysis, re-collect the sample.
2. Direct Interference: After hemolysis, blood chemical constituents that are present in high concentrations within RBCs will show increased concentrations in the serum. Therefore, avoid using hemolyzed specimens for the determination of chemical constituents that are abundant in RBCs, such as potassium, ALP, AST, bilirubin, CK, LD, and ACP.
Since centrifugal samples—beyond just blood—generally have corrosive properties, even ordinary tap water can corrode aluminum rotors. When stress corrosion and chemical corrosion occur simultaneously, a phenomenon known as dual-corrosion fatigue may arise, potentially reducing the rotor’s actual service life below its guaranteed lifespan. If certain rotors are left unused for extended periods, their internal metal structures may undergo changes; therefore, such rotors should be operated at reduced speeds. The manufacturer provides a warranty period of 5 years. Consequently, users are required to diligently record the number of operating cycles and the cumulative operating time for each rotor. Thanks to advances in technology, modern centrifuges are equipped with imbalance detection and protection systems that automatically perform safety safeguards and alarm functions, so users need not worry excessively when using them.
Predicting failure by monitoring operational and physical indicators is a better approach. In centrifuges, the cover spring mechanism might loosen, bearings could wear out, and even with proper lubrication, the ear shaft might stop functioning. Any one of these issues is cause for concern, but taken together, they all point to impending failure.
Please place the “Operating Instructions” in an easily accessible location near the centrifuge equipment, ensuring that all users can promptly read this manual. Warning: Proper transportation, storage, positioning, installation, as well as careful operation and maintenance are essential for the centrifuge’s normal and safe operation. During transportation and storage, ensure that the centrifuge is protected from physical impacts. Additionally, the instrument must be kept away from rain and should not be stored in environments with excessively high or low temperatures or high humidity levels. For specific operating instructions, please read the manual carefully.