Reasons Why a Low-Temperature Centrifuge Cannot Cool

Release time:

2014-02-24


Overview:

There are two possible reasons why the temperature is set to normal but no cooling occurs: (1) Rule out a malfunction of the temperature control panel; (2) Check for a sensor failure.

There are two possible reasons why the unit can normally set the temperature but fails to cool:

(1) After ruling out a fault in the temperature control panel, we suspected a problem with the refrigeration system. Users had reported that the centrifuge’s cooling performance had been poor in recent days. Upon inspection, we found that the refrigerant level in the air compressor was too low, resulting in inadequate cooling. We replenished the compressor with refrigerant, and the fault was successfully resolved.

(2) Sensor failure. Temperature is an important parameter for refrigerated centrifuges, as certain samples—such as cells and proteins—can be damaged at high temperatures. During high-speed operation, the centrifuge generates heat; therefore, it is essential to use a refrigeration system to cool the centrifuged samples and maintain the centrifugation environment within a specific temperature range (e.g., 1–10°C). After opening the chassis cover and checking the voltage on the temperature control panel, we found that the 12 V and 6 V voltages were normal. We suspected that the temperature-control integrated circuit (ICL7107CPL L9646) might be malfunctioning. Fortunately, we had an identical temperature-control IC (ICL7107CPL 58840) salvaged from a decommissioned old machine. After replacing the faulty chip, the centrifuge’s malfunction was resolved, and the machine returned to normal operation. If sensor failure causes inaccurate temperature readings and leads to erroneous data that triggers the refrigeration system to switch on and off randomly, the sensor encased around the exterior of the centrifuge body must be replaced.