Common Vibration Detection Methods and Device Design for Centrifuges

Release time:

2014-03-08


Overview:

The commonly used vibration detection method for centrifuges is to measure the amplitude of the centrifuge's vibrations.

The usual practice is to install limit switches or proximity switches at specific locations on the centrifuge. When the vibration amplitude of the centrifuge reaches a certain level, the limit switch is activated or the proximity switch detects a signal, thereby cutting off the power supply to the drive motor and initiating braking. Under the combined action of gravity and braking force, the rotational speed and vibration amplitude of the rotating body gradually decrease, thus protecting the centrifuge from damage and preventing accidents. Since the radial impact force experienced by the centrifuge varies depending on its rotational speed at the same vibration amplitude—where the radial impact force caused by dynamic imbalance is proportional to the square of the rotational speed—it is insufficient to rely solely on the centrifuge’s vibration amplitude as a protection criterion.

The dynamic imbalance protection for centrifuges should use the radial impact force acting on the shaft system as the vibration protection criterion. According to Newton’s Second Law, the force applied to an object is directly proportional to its acceleration. Therefore, the impact force can be measured by detecting the acceleration of the centrifuge’s body. The centrifuge’s automatic controller can then regulate the centrifuge’s operation based on the body’s acceleration and rotational speed, thereby reducing impacts and preventing accidents.
 
Design of a Vibration Detection and Protection Device for Centrifuges

By changing the capacitance values of capacitors Cy and Cx connected to pins 11 and 12, the bandwidth of the ADXI 210 can be adjusted between 0.01 Hz and 5 kHz. The typical noise of the ADXI 210 is 500 µg/H. Under bandwidth conditions below 60 Hz, the sensor’s resolution can reach 5 mg. The internal filtering resistor of the acceleration sensor is 32 kΩ. The equation for the acceleration sensor at a 3 dB bandwidth is: 1. — [2πRXC—(x,y)].
Among them, R = 32 kΩ or, simplified: F = 3 dB, Y
Since the filtering resistor (Rfih) inside the accelerometer may have a deviation of up to 25%, the bandwidth calculated according to the above formula will also exhibit a corresponding deviation. The filter in the accelerometer helps reduce noise in the measured acceleration and improves measurement resolution.

Since the centrifuge’s rotational speed is below 1,500 rpm, its impact acceleration is a positive function of the centrifuge’s rotational speed, and its frequency should therefore be less than 30 Hz. According to Shannon’s sampling theorem, to accurately measure the centrifuge’s impact acceleration, the accelerometer’s bandwidth must exceed 60 Hz. Taking into account the tolerance of the internal filtering resistors in the sensor, the accelerometer’s bandwidth should even exceed 100 Hz. To minimize errors caused by IX2M (Duty Cycle Modulation), the filter’s bandwidth should be less than one-tenth of the DCM frequency; consequently, the DCM frequency should exceed 1 kHz.