A deeper interpretation of low-speed centrifuges
Release time:
2015-11-24
Overview:
Low-speed centrifuge
Low-speed centrifuge A literal interpretation often leads one to think of a centrifuge with relatively low speed. However, today we’d like to clarify that assuming this straightforwardly could actually be misleading. After all, low-speed centrifuges aren’t necessarily characterized solely by their low rotational speed; rather, the key factor is the slow rotation of the rotor itself, which ties directly to rotor-related considerations. Additionally, the nature of the experimental samples or specific experimental requirements also play a crucial role in determining whether a centrifuge is classified as low-speed.
When the rotor of a low-speed centrifuge spins at high speed, air friction generates heat, causing the rotor's temperature to rise and, consequently, the temperature of the samples in the tubes to increase as well. Since biological samples are highly sensitive to temperature, it is generally required that the temperature during centrifugation experiments be maintained at 4℃. Without intelligent technology controlled by a computer, it would be impossible to achieve an accuracy of ±1℃. Therefore, refrigeration units are used to cool the centrifuge chamber, thereby cooling both the rotor and the sample temperatures. However, measuring the actual temperature of a rotating rotor is extremely challenging. Both domestically and internationally, temperature sensors are typically placed near the rotor at the bottom of the centrifuge chamber to indirectly measure the rotor’s temperature. In this case, the sample temperature Tsample equals the chamber temperature Tchamber plus a compensation term Tcompensation. The compensation term Tcompensation, in turn, depends on the rotor’s type, its rotational speed, and the duration of operation. For low-speed centrifuges, voltage fluctuations must comply with the national standard, staying within ±10%. If voltage fluctuations exceed this range, it is recommended to use a voltage stabilizer. Additionally, the power wiring must meet the required specifications: when using three-phase power, the neutral wire and the ground wire should be kept separate. More importantly, whether it’s three-phase industrial power or ordinary single-phase power, the grounding wire must be reliable to prevent electric leakage.
Install Low-speed centrifuge When installing, be sure to accurately identify the installation location. During installation, keep the centrifuge lid as open as possible. Use a small spirit level to check for horizontal alignment on the main shaft, then adjust the leveling bolts next to the four casters to achieve proper balance. One particularly important point to note is that none of the four feet should be unsupported. For the centrifuge, to ensure its normal operation in the future, the floor must be solid and free from any depressions. The ground at the installation and fixing location should be both smooth and stable.
During operation, the low-speed centrifuge can accumulate frost inside its chamber due to moisture in the air. Once the machine is turned off, this frost melts into water. Since most centrifuges are not equipped with drain ports, over time the amount of accumulated water in the centrifuge chamber will keep increasing. Therefore, during use, we need to regularly clean out the accumulated water.
A typical manifestation of analog technology is the use of knobs to select operational parameters (such as rotational speed, temperature, and time), with the needle on the dial serving as the display for these values. Because the selection of parameters and the reading of data are heavily influenced by human factors—both from the operator and the person taking the readings—the control accuracy is poor. Moreover, this process must be repeated every time, leading to poor repeatability. Therefore, although some centrifuges feature digital displays (e.g., using knobs to set values while the digits show the readings), even though there’s an improvement in readability, the underlying principle for value acquisition remains unchanged and thus still falls under analog technology. In contrast, a typical digital display system features a user-friendly interface, keyboard operation, digital numerical display, and fully computer-controlled programmable operations. At its core lies intelligent control, which is realized entirely by computers. Low-speed centrifuge All parameters required for operation—such as rotational speed, temperature, time, and acceleration/deceleration rates—can be entered via keyboard input and displayed digitally.
When the rotor of a low-speed centrifuge spins at high speed, air friction generates heat, causing the rotor's temperature to rise and, consequently, the temperature of the samples in the tubes to increase as well. Since biological samples are highly sensitive to temperature, it is generally required that the temperature during centrifugation experiments be maintained at 4℃. Without intelligent technology controlled by a computer, it would be impossible to achieve an accuracy of ±1℃. Therefore, refrigeration units are used to cool the centrifuge chamber, thereby cooling both the rotor and the sample temperatures. However, measuring the actual temperature of a rotating rotor is extremely challenging. Both domestically and internationally, temperature sensors are typically placed near the rotor at the bottom of the centrifuge chamber to indirectly measure the rotor’s temperature. In this case, the sample temperature Tsample equals the chamber temperature Tchamber plus a compensation term Tcompensation. The compensation term Tcompensation, in turn, depends on the rotor’s type, its rotational speed, and the duration of operation. For low-speed centrifuges, voltage fluctuations must comply with the national standard, staying within ±10%. If voltage fluctuations exceed this range, it is recommended to use a voltage stabilizer. Additionally, the power wiring must meet the required specifications: when using three-phase power, the neutral wire and the ground wire should be kept separate. More importantly, whether it’s three-phase industrial power or ordinary single-phase power, the grounding wire must be reliable to prevent electric leakage.
Install Low-speed centrifuge When installing, be sure to accurately identify the installation location. During installation, keep the centrifuge lid as open as possible. Use a small spirit level to check for horizontal alignment on the main shaft, then adjust the leveling bolts next to the four casters to achieve proper balance. One particularly important point to note is that none of the four feet should be unsupported. For the centrifuge, to ensure its normal operation in the future, the floor must be solid and free from any depressions. The ground at the installation and fixing location should be both smooth and stable.
During operation, the low-speed centrifuge can accumulate frost inside its chamber due to moisture in the air. Once the machine is turned off, this frost melts into water. Since most centrifuges are not equipped with drain ports, over time the amount of accumulated water in the centrifuge chamber will keep increasing. Therefore, during use, we need to regularly clean out the accumulated water.
A typical manifestation of analog technology is the use of knobs to select operational parameters (such as rotational speed, temperature, and time), with the needle on the dial serving as the display for these values. Because the selection of parameters and the reading of data are heavily influenced by human factors—both from the operator and the person taking the readings—the control accuracy is poor. Moreover, this process must be repeated every time, leading to poor repeatability. Therefore, although some centrifuges feature digital displays (e.g., using knobs to set values while the digits show the readings), even though there’s an improvement in readability, the underlying principle for value acquisition remains unchanged and thus still falls under analog technology. In contrast, a typical digital display system features a user-friendly interface, keyboard operation, digital numerical display, and fully computer-controlled programmable operations. At its core lies intelligent control, which is realized entirely by computers. Low-speed centrifuge All parameters required for operation—such as rotational speed, temperature, time, and acceleration/deceleration rates—can be entered via keyboard input and displayed digitally.