Centrifuge Calculation Formula in RNA Extraction Process

Release time:

2014-03-13


Overview:

During RNA extraction, a centrifuge is used. By default, the centrifuge’s centrifugal force is expressed in rpm. However, the instructions for Trizol typically specify centrifugation requirements in terms of relative centrifugal force (RCF).

During RNA extraction, a centrifuge is used. By default, the centrifuge’s centrifugal force is expressed in rpm. However, the instructions for Trizol typically specify centrifugation requirements in terms of relative centrifugal force (RCF).

RPM (revolutions per minute) refers to the number of revolutions a centrifuge makes per minute (unit: revolutions per minute); RCF (relative centrifugal force) is the relative centrifugal force, expressed as a multiple of Earth's gravitational acceleration, typically denoted by ×g.

r is the radius of the centrifuge rotor (unit: meters).

The conversion relationship between the two is as follows:

rcf = 0.0011 × r × rpm²

rpm = (909 × rcf ÷ r)^0.5

Taking the Eppendorf centrifuge 5810r, F45-30-11 rotor as an example: max RPM = 1400 rpm, r = 9.5 cm. The Trizol protocol requires an RCF of 12,000 × g when extracting RNA.

Then the required RPM = (12000 × 909 ÷ 0.095)^0.5 = 10,715 rpm