Common Troubleshooting for Laboratory Freezer Centrifuges That Do Not Display Data and Fail to Operate Normally

Release time:

2013-07-03


Overview:

The centrifuge can run, but its maximum speed is only 2500 r/min, which does not meet the operational requirements. Repair and troubleshooting: The adjustable speed resistor WW was inspected and found to be undamaged. The gate trigger voltages Ug of thyristors SCR1 and SCR2 were measured; the Ug voltage, which is adjustable at 7V, is within the normal range.

1 Failure One
The centrifuge’s power indicator light and instruments are normal, but when the start button is pressed, the centrifuge does not run.
Maintenance and Troubleshooting: The control circuit was inspected and no faults were found. After replacing the centrifuge’s carbon brushes, the machine resumed normal operation. This phenomenon is frequently encountered and is caused by worn-out carbon brushes.
2 Fault No. 2
The centrifuge can run, but its maximum speed is limited to only 2,500 rpm, failing to meet the operational requirements. Repair and troubleshooting: The adjustable speed resistor WW was inspected and found to be undamaged. The gate trigger voltages Ug of thyristors SCR1 and SCR2 were measured; the Ug voltage of 7V is adjustable and within the normal range. The forward conduction and reverse blocking characteristics of both SCR1 and SCR2 appeared normal on the surface. However, after replacing SCR2, the centrifuge’s rotational speed could be adjusted up to 3,200 rpm. The analysis indicates that the abnormality was caused by parameter variations in SCR2.
3 Failure Three
Transformer B has burned out, causing the entire machine to fail. Repair and troubleshooting: Transformer B’s primary winding is rated at 380V, and its secondary winding provides three output voltages: 220V, 150V, and 160V. The contacts of CJ1 and CJ2 are connected in series across the secondary winding (0–150V) of Transformer B; CJ1 and CJ2 are interlocked contacts. After Transformer B burned out, the contacts of CJ1 and CJ2 were inspected. It was found that the contacts of CJ1 had fused together, resulting in damage to CJ1 itself. After replacing CJ1 and the transformer, the centrifuge returned to normal operation. Analysis indicates that the cause of the failure was an excessively high inrush current, which led to the burnout of CJ1.
4 Failure Four
The start-up speed indicator shows 3,000 rpm; the centrifuge’s rotational speed cannot be lowered.
Handling and Repair: The gate trigger voltage Ug of the thyristor was measured at 7V and is fixed—non-adjustable. Upon inspection, both the control-stage component T5 and the light-emitting diode GD in the preceding stage were found to be functioning properly. Further inspection led to the discovery that the emitter resistor R35 of transistor T3 had changed from 33n to 180Ω. After replacing R35, the fault condition was eliminated.