Analysis of the Causes of Blood Bag Damage During Centrifugation in Blood Bank Centrifuges
Release time:
2013-05-10
Overview:
1. Dead Space: The "dead space" is the primary cause of centrifugal damage. During centrifugation, blood bags are subjected to tremendous centrifugal forces—equivalent to weights ranging from hundreds of kilograms to over a ton per bag. If the blood bag were in full contact with the centrifuge cup, the cup would bear the full brunt of this enormous force, thus preventing any damage to the blood bag. However, if a dead space exists between the blood bag and the centrifuge cup, the thin sheet—approximately 0.4 mm thick—cannot withstand the force exerted by the blood drawn into the dead space, leading to deformation (wart-like protrusions) and eventual rupture.
2. Catheter Exposure: Due to insufficient placement of the blood bag in the centrifuge cup, the catheter and its protective sheath may end up exposed at a high position near the top edge of the centrifuge cup’s opening. Alternatively, if the catheter hangs outside the centrifuge cup during centrifugation, it could be pulled out of the cup without any obstruction. As a result, the catheter may break, and the Clicktip could damage the protective sheath. Shanghai Luxiangyi Centrifuge would like to remind you that this situation could lead to extremely dangerous incidents, such as the catheter becoming entangled around the centrifuge’s rotor shaft.