Centrifugation Requirements for Serum Separator Tube Coagulation-Inducing Blood Collection Tubes

Release time:

2012-03-21


Overview:

To prepare high-quality serum samples using separation-gel-coated tubes, two critical steps are the complete clotting of the blood and proper centrifugation conditions. For centrifugation, a horizontal rotor in a benchtop centrifuge is required.

Use serum-separating gel tubes to prepare high-quality serum specimens. Two critical steps are ensuring complete blood clotting and proper centrifugation conditions. For centrifugation, a horizontal rotor in a benchtop centrifuge is required.

The specific operational steps are as follows: Immediately after blood collection, gently invert the blood collection tube 4–5 times to thoroughly mix the sample. Allow the sample to fully clot for 30 minutes. Use a centrifuge with a radius of 8 cm and maintain a centrifugal speed of 3,500–4,000 rpm for 10 minutes.
The serum is completely separated from the blood clot by the separating gel, and the serum sample can either be directly loaded onto the instrument for testing or transferred into the instrument’s dedicated test cups.

Only when this condition is met can high-quality serum specimens be prepared, indicating that the separation gel performs well. If the centrifugation speed is too low, the force acting on the separation gel will be insufficient, leading to poor separation of the gel layer or even incomplete blood clotting before centrifugation. As a result, fibrin clots may remain suspended in the serum or within the gel layer, potentially causing hemolysis. Except for emergency cases, it is best to centrifuge routine biochemical samples only after the blood has completely clotted for optimal results. Due to lack of experience, this phenomenon frequently occurs when laboratories first begin using separation-gel vacuum tubes. Moreover, if fibrin strands remain in the serum, they can easily clog the blood-collection needles of automated analyzers.

Currently, the quality of many domestically produced separation gels has reached or even surpassed international standards. Usage in numerous hospitals has shown that the separation time and separation performance of these gels are virtually indistinguishable from those of comparable products from companies such as BD in the U.S. We conducted a study to investigate whether using domestically produced (Youli) separation-gel plastic tubes affects the measured values for more than 30 biochemical and immunological assays. We used ordinary glass tubes as a control and performed comparative experiments against similar products from BD in the U.S. The results demonstrated excellent comparability between the two types of tubes, with no statistically significant differences observed (both p > 0.05).