Using a laboratory centrifuge requires familiarity with the functions of each component.
Release time:
2012-02-27
Overview:
1 Precise installation
(1) Power supply voltage: Voltage fluctuations must comply with the national standard, which allows for variations within 10%. Otherwise, centrifuges from certain manufacturers may fail to operate properly. If voltage fluctuations exceed this limit, it is recommended to use a voltage stabilizer to ensure compatibility with the power grid.
(2) The grounding wire must be secure: The wiring for the room’s power supply must meet the required standards. In China, the power system uses a three-phase four-wire configuration. The three phases refer to the three phases used in industrial electricity. When using three-phase power, the neutral wire and the ground wire must be kept separate. More importantly, whether it’s three-phase industrial power or ordinary single-phase power, the ground wire must be reliably connected—this is crucial to prevent electric leakage. Under no circumstances should the ground wire be connected to heating pipes or water pipes.
(3) The floor must be level and stable: For large centrifuges, when moving the floor at the installation site, ensure it is stable to guarantee that the centrifuge rotor operates normally without sinking or deforming. The floor at the installation location should be both level and stable.
(4) Adjust the level: Locate a precise, small-leveling gauge. With the centrifuge lid open, find the level on the spindle. Adjust the level by turning the adjusting bolts under the four casters. Pay special attention to ensure that one of the casters does not lose contact with the ground.
2. Load samples and find equilibrium.
When designing and manufacturing centrifuges, manufacturers have already taken corrective measures through dynamic balancing tests to address the imbalance caused by machining tolerances in the rotor. However, centrifuges typically have a permissible limit for sample loading imbalance. To compete with other manufacturers on the maximum allowable imbalance level, centrifuge manufacturers tend to specify the highest possible value. While the centrifuge can indeed operate within this higher limit, the resulting unbalanced forces will suddenly impact SKF bearings and support structures at a frequency of n times per minute, thereby causing damage to the centrifuge. Therefore, for users of these expensive centrifuges, it is highly advisable to ensure proper balancing before starting the centrifuge—doing so will significantly extend the machine's service life.
3. Remove accumulated water from the centrifuge chamber.
When the centrifuge is running, refrigeration causes moisture in the air to condense into frost inside the centrifuge chamber. After the machine is stopped, this frost melts into water. Most low-speed, high-capacity centrifuges available internationally lack drain holes, leading to an accumulation of water in the centrifuge chamber. In such cases, users should remove the rotor themselves and carefully drain the accumulated water. When reassembling the rotor, be sure to install it correctly to prevent accidents or malfunctions.
4. Aluminum alloys cannot be corroded.
Centrifuge rotors are typically made of aluminum alloy. Once corroded, their strength diminishes, increasing the risk of accidents and malfunctions. Aluminum alloys are susceptible to liquid corrosion; after cleaning, be sure to dry them thoroughly using a hairdryer or by inverting them for a while until they’re completely dry before use. Some blood banks’ centrifuges lack the six plastic holders that should normally be included in the six aluminum cups. These plastic holders are designed to hold blood bags—never place blood bags directly into the aluminum cups. Also, carefully inspect test tubes for cracks. If a test tube—especially one that has been used repeatedly—shows even the slightest crack, never use it under any circumstances. Otherwise, during operation, a cracked test tube could cause an imbalance during rotor rotation, potentially leading to catastrophic incidents such as shaft breakage.
5. The Broken-Axle Incident
If the rotor is not properly balanced or if the sample loading is excessively unbalanced, the centrifuge will be started. Although modern entry-level centrifuges are equipped with imbalance protection—meaning that when the imbalance exceeds a certain threshold, the machine will automatically cut power and stop spinning—by the time such a situation arises, it’s already too late, and a serious accident could occur. One of the most severe accidents is shaft failure: during high-speed rotation, a shaft may suddenly break, leaving the centrifuge rotor unsupported and causing it to spin wildly and crash around inside the centrifuge chamber. This can result in the entire centrifuge rotating through an angle of up to 270 degrees. At this point, if there are personnel or equipment nearby, it could lead to injuries or even fatalities among staff and catastrophic damage to the equipment.