Simple analysis of the temperature settings for the sample in the centrifuge chamber.
Release time:
2022-08-23
Overview:
I’ve written many technical articles about centrifuges, but not a single one has ever focused on the temperature inside the centrifuge chamber. So today, let’s talk about that.
Take the refrigerated centrifuge as an example: Some experimental studies have found that, at the same temperature setting, the rotational speed affects the deviation between the sample temperature and the set temperature—in a directly proportional relationship. In other words, the higher the rotational speed, the greater the deviation.
Detailed explanation as follows:
As the rotational speed increases, the deviation between the actual air temperature inside the chamber and the set temperature also becomes greater. At different rotational speeds, by setting different chamber temperatures for temperature compensation, it is possible to maintain the temperature of the sample reagents around 4℃ throughout the specified centrifugation test duration. Thus, we have preliminarily developed a method for keeping the sample reagents inside the centrifuge rotor at the temperature required for the centrifugation test. The experimental data and analysis results provide valuable references and guiding principles for future product development and practical application.
Detailed experimental protocol for temperature measurement of sample reagents within the cavity:
Since it is extremely difficult to measure the actual temperature of the sample reagent inside a rotating rotor, both domestically and internationally, temperature sensors are typically placed near the bottom of the centrifuge chamber, close to the rotor. By controlling the temperature within the centrifuge chamber, the temperature of the sample reagent inside the rotor can be indirectly regulated. This experimental approach also adopts this method. The relationship between the sample temperature and the chamber temperature is as follows:
T sample = T cavity + T compensation, where:
The repayment is related to the type of rotor, its rotational speed, and the duration of operation.
Test conditions:
The working environment temperature is 26°C, and the centrifuge is powered by a 60Hz voltage-stabilized power supply.
The centrifuge rotor’s minimum rotational speed is 2,000 r/min, and its maximum rotational speed is 13,000 r/min.
Select three different rotational speeds as test speeds: 2000 r/min, 10,000 r/min, and 13,000 r/min.
Considering also that typical plasma and cell separation assays are completed within 30 minutes, we selected seven time points—5, 10, 15, 20, 25, 30, and 35 minutes—to measure the temperature of the sample reagents inside the rotor.
However, the refrigerated centrifuges sold by Shanghai Luxiangyi Centrifuge Instrument Co., Ltd.—regardless of their capacity, rotational speed, or whether they are horizontal or vertical—feature high-performance imported compressors and use the fluorine-free refrigerant R404a, meeting environmental protection standards and delivering excellent cooling performance. Their temperature control accuracy reaches ±1℃, making them widely applicable in fields such as clinical medicine, biochemistry, genetic engineering, and immunology. They are particularly suitable for the separation and purification of radioimmunoassay samples, water treatment solutions, biochemical substances, biopharmaceutical products, and blood derivatives.