The safety design of low-speed centrifuges is carefully considered.
Release time:
2020-11-19
Overview:
Low-speed centrifuge It features direct motor drive and a special vibration-damping system. The rotor is precision-cast from high-strength materials, ensuring easy assembly and disassembly. Equipped with intelligent control, the device offers fully digital displays for both rotational speed and timing, guaranteeing precise control. A single rotor can be easily interchanged with various types of tube racks, meeting the diverse needs of different users.
The safety design of low-speed centrifuges is carefully considered:
1. The door-lock protection system features a dual-layer locking mechanism—mechanical and electronic. The centrifuge can only start after the door lock is closed, and the lid can only be opened after the motor has come to a complete stop.
2. Enhancing the centrifuge’s inherent ability to withstand imbalance not only improves its robustness but also simplifies user operation. When using the rotor, there is no need for strict balancing—indeed, visual balancing alone is often sufficient. Typically, small centrifuges mount the entire motor on vibration-damping elastic supports, such as rubber pads. High-speed centrifuges often employ flexible shafts designed to absorb vibrations. By suspending the entire motor from rubber mounts and placing the entire shaft within a vibration-damping buffer container, these designs overcome the drawbacks of flexible shafts—namely, their tendency to bend and break—and thus offer greater structural integrity, enhanced resistance to imbalance, and faster acceleration and deceleration.
3. The rotor recognition system can automatically identify rotors and limit their maximum rotational speeds, effectively preventing user errors. The basic principle behind rotor recognition is to assign a unique code to each rotor—or to embed a specific number of tiny magnets at the base of each rotor—and then use Hall sensors or photoelectric refractive ring sensors for detection. More sophisticated methods include ultrasonic and radio-frequency wireless rotor identification, which allow rotors to be recognized without even needing to be rotated. These systems can even detect whether the rotor lid is properly closed.
4. The centrifuge chamber and lid are made of robust armor steel to prevent rotor fragments from flying off. It is crucial to detect rotor imbalance, as rotor imbalance can cause the motor shaft to oscillate, shortening the centrifuge’s lifespan. In severe cases, it can even lead to breakage of the motor shaft or rotor explosion.
Low-speed centrifuge Common Fault Analysis:
After the low-speed centrifuge is powered on, its control panel displays normally, and the cooling fan operates properly. However, when the start button is pressed, the intermediate relay activates, yet the centrifuge fails to start. Based on the observed fault symptoms, the power supply and control sections appear to be functioning correctly; therefore, the fault must lie in the wiring between the control board and the motor, as well as in the carbon brushes and the motor itself. Upon opening the centrifuge, a multimeter was used to verify that the wiring from the control board to the carbon brushes is conductive. However, when measuring the continuity from the carbon brushes to the motor rotor, no conductivity was detected. Upon closer inspection, it was found that the carbon brushes had poor contact with the motor rotor. After removing the carbon brushes, it became evident that they had become excessively worn due to prolonged use, leaving them too short. Replacing the carbon brushes with new ones successfully resolved the issue. This type of failure is a common problem in centrifuges, so it’s important to regularly replace the carbon brushes to prevent similar faults from occurring.
After powering on the low-speed centrifuge, the control panel displays normally, and the cooling fan operates properly. However, when the start button is pressed, the intermediate relay fails to engage, and the centrifuge does not start. Based on the observed fault symptoms, the power supply appears to be functioning correctly; therefore, the fault must lie in the control board, the wiring between the control board and the motor, the carbon brushes, or the motor itself. Upon opening the centrifuge, a multimeter was used to verify that the wiring from the control board to the carbon brushes is conductive. Further testing confirmed that the wiring from the carbon brushes to the motor rotor is also conductive, and the motor windings were found to be normal.
After powering on, using a multimeter to measure the wires connecting the control board to the motor revealed no voltage output. This indicates that the fault lies in the control board itself. Specifically, the intermediate relay is not functioning, so the fault must be in the control circuit of the intermediate relay. By systematically troubleshooting each component, we identified that the C9013 transistor on the control board had been damaged. We replaced it with a new C9013 transistor, and the fault was successfully resolved.