What operational details should be noted when using a low-speed centrifuge?
Release time:
2021-10-08
Overview:
Low-speed centrifuge This is a microcomputer-controlled, desktop low-speed centrifuge with a steel body and a powder-coated surface, offering excellent rigidity and high strength. It features a novel design and an aesthetically pleasing appearance, along with advantages such as low noise, minimal temperature rise, and high safety and reliability. Equipped with a brushless motor drive, microcomputer control, door-lock protection, and automatic rotor balancing, this centrifuge ensures safer, simpler, and more reliable operation. It is widely used in research laboratories and production facilities across fields including radioimmunoassay, biochemistry, and pharmaceuticals for the separation of particles with varying densities.
The low-speed centrifuge features a reasonably designed structure that effectively eliminates sanitary dead zones. Its structural components have smoothly rounded transitions, and their surfaces are polished. All external structural parts, fasteners, and the rotor drum are made of stainless steel. The housing adopts a flip-top design, allowing for thorough cleaning of the space between the outer casing and the rotor drum. The centrifuge can be started using a frequency converter, ensuring smooth startup and adjustable separation factors. It is equipped with a non-contact energy-regenerative braking system, a static-electricity-resistant belt drive, and a fully enclosed body structure. Sealing elements are made of silicone rubber or fluororubber. This design meets stringent sealing and explosion-proof requirements, making it suitable for use in environments involving toxic, flammable, and explosive substances.
A centrifuge utilizes the principle of centrifugal sedimentation to separate, concentrate, or purify cells (particles) with different densities in a solution under the action of centrifugal force. Centrifuge tubes containing equal volumes of sample are placed symmetrically in the rotor’s tube slots. After the instrument is started, the motor drives the rotor to spin at high speed, generating relative centrifugal force (RCF) that causes cells (particles) of varying densities in the sample to separate.
Detailed precautions for using low-speed centrifuges:
1. Open the door
Press the stop button to open the door cover (in the event of a power outage, you can use the side mechanical lock handle to open the door cover).
2. Rotor Installation
Wipe the drive shaft inside the centrifuge chamber and the inner bore of the rotor with a clean, soft cloth. Hold the rotor horizontally with both hands, align the slots with the shaft pins, and carefully insert it vertically, ensuring that the rotor slots fit snugly over the drive shaft pins. Then, install the washer and gently tighten it using a screwdriver.
Attention: When using a plastic rotor, please do not tighten it too much to avoid damaging the rotor!
3. Place it into the test tube.
Place centrifuge tubes containing equal amounts of sample solution symmetrically into the rotor’s tube holes to ensure balanced operation of the rotor.
4. Close the door
Close and secure the door cover, then verify that it’s properly locked (after closing, gently apply light pressure to the door cover—if it doesn’t budge, it’s locked). If the door cover isn’t fully closed, pressing the start button will cause the time window to display an E3 alarm.
5. Parameter Settings
a) Plug in the power cord and turn on the power switch.
b) Rotational Speed Setting: According to Table 1 and the required separation speed for the solution, press the “+” or “–” keys in the rotational speed window to set the desired centrifugal speed.
Note: The set rotational speed must not exceed the rotor’s ***maximum rotational speed.
c) Time setting: In the time window, use the “+” and “–” keys to set the required time according to the centrifugation duration.
6. Start and run
a) Check whether the “speed setting” and “centrifugation time” are accurate and correct.
b) Press the “Start” button; the start indicator light will illuminate, and the rotational speed will begin to increase until it reaches the preset speed, after which it will run at a constant speed. The timing is displayed in countdown mode.
c) When it is necessary to change the separation parameters (rotational speed and time) during operation, press the “+” and “-” keys below the rotational speed window to adjust the rotational speed parameter; similarly, press the “+” and “-” keys below the time window to adjust the time parameter. The instrument will automatically switch to operating with the newly adjusted parameters.
d) When the time display reads “0,” the centrifuge automatically cuts off power to the drive motor, causing the rotational speed to decrease. Once the rotational speed reaches “0,” the start indicator light turns off, the stop indicator light turns on, and the instrument automatically emits a beep to notify the operator.
e) If it is necessary to stop the machine midway, press the stop button, and the centrifuge will automatically cut off the power supply to the drive motor.
7. After operation is complete, when not in use, please turn off the power switch and unplug the power cord.