Blood separation and purification rely on a low-speed centrifuge.
Release time:
2020-12-18
Overview:
Low-speed centrifuge What aspects does the system consist of?
Low-speed centrifuges have rotational speeds below 10,000 rpm. Common application industries include textiles, chemical processing, and environmental protection, among others.
However, in recent years, low-speed centrifuges have rapidly gained popularity in the pharmaceutical industry.
Many pharmaceutical companies need to separate mother liquors, and in such cases, low-speed centrifuges become crucial.
Meanwhile, due to the stringent regulations in the pharmaceutical industry, pharmaceutical centrifuges are predominantly of the flat-sealed type. To minimize potential contamination or damage and enhance hygiene, the parts that come into contact with materials are made of stainless steel or entirely constructed from stainless steel. The entire machine has no sanitary dead spots, making it clean and easy to use. This type of centrifuge has now become widespread throughout the pharmaceutical industry.
In addition, blood banks or hospital hematology departments also use low-speed centrifuges operating at around 3,000 rpm to separate blood.
When many patients go to the hospital, doctors usually draw blood and collect it in vacuum blood collection tubes, which are then taken to the laboratory for testing. Those who pay close attention may notice that most doctors place these vacuum blood collection tubes into a centrifuge for centrifugation. The following describes the process of using a centrifuge in a hospital’s laboratory.

First, after the doctor receives the blood collection tubes, they will place them on a rack designed for vacuum blood collection tubes. Once a sufficient number of tubes have been collected, they’ll be transferred to a centrifuge for spinning. It’s crucial to use an even number of tubes, as a single tube cannot be centrifuged. Centrifuges require balance during operation; although some centrifuges are equipped with automatic balancing systems, it’s still impossible to centrifuge just one tube. Rather, as long as the weight difference among the tubes isn’t too significant, the centrifuge can maintain normal operation. During centrifugation, the centrifuge generates tremendous centrifugal force. If the system becomes unbalanced, it could lead to accidents. All Luxiangyi centrifuges are equipped with protective devices that prevent them from operating normally under conditions of imbalance.
But you still need to pay attention:
1. For the centrifugal separation of serum from specimens, the relative centrifugal force (RCF) of the centrifuge should be set at 1000g to 1200g, and the centrifugation time should be 5 to 10 minutes. Increasing either the relative centrifugal force or the centrifugation time can easily lead to hemolysis.
II. Handling of Hemolyzed Specimens: Hemolysis can lead to the release of various components from RBCs, thereby interfering with assay methods. Hemoglobin (Hb) levels ≥0.2 g/L indicate visible hemolysis.
1. Indirect Interference: The release of hemoglobin can cause indirect interference, as hemoglobin exhibits light absorption at wavelengths of 431 nm and 555 nm. If the analyte being measured is analyzed colorimetrically using wavelengths close to these, it may lead to falsely elevated results. Treatment methods: For mild hemolysis, perform a serum blank simultaneously; for severe hemolysis, re-collect the sample.
2. Direct Interference: After hemolysis, blood chemical constituents that are present in high concentrations within RBCs will show increased concentrations in the serum. Therefore, avoid using hemolyzed specimens for determining the levels of chemical constituents that are abundant in RBCs, such as potassium, ALP, AST, bilirubin, CK, LD, and ACP.
For a routine biochemical test, you’ll need to separate the serum. This is a low-speed centrifuge, operating at a speed of around 3,000 to 5,000 rpm. Typically, the process takes no more than 3 minutes.
Serum separation typically involves centrifugation on the day of blood collection; if stored appropriately, it can be kept for up to one day. Hemolysis usually occurs when blood is transferred from the syringe into the centrifuge tube too quickly and under excessive pressure after collection, or when the centrifugation speed is too high—hemolysis tends to occur at speeds above 8,000 rpm.
The above is a brief explanation of how low-speed centrifuges are used for blood separation. If you’d like to learn more about centrifuge applications, please stay tuned to Shanghai Luxiangyi Centrifuge Instrument Co., Ltd.—the company regularly updates its blog with handy little tips and insights on centrifuge-related topics!