The service life of a desktop centrifuge is related to its maintenance.

Release time:

2017-03-16


Overview:

How long a benchtop centrifuge will last depends largely on how carefully it’s treated. With proper maintenance, it can serve you for a long time. Many customers find that some benchtop centrifuges need repairs every few months, while others can run smoothly for a year and a half or even longer without any issues at all. Therefore, to ensure your benchtop centrifuge has a long lifespan, it’s crucial to give it good care and maintenance.
   Desktop centrifuge How long a device will last depends largely on how carefully people treat it. If you take good care of it and maintain it properly, it can serve you for a long time. Many customers using desktop centrifuges find that some need repairs every few months, while others don’t encounter any issues for a year and a half or even longer. Therefore, to ensure the longevity of your desktop centrifuge, proper maintenance is essential.
  Key points for using a desktop centrifuge:
  The benchtop centrifuge is a commonly used product type, widely employed in fields such as chemical engineering, food processing, pharmaceuticals, mineral processing, coal mining, water treatment, and shipbuilding. Users need to master the key points of operating a benchtop centrifuge, which will help them make better use of this equipment. Let’s take a closer look at the specific details below.
  1. If the main power indicator light of the centrifuge is on, you should check the degree of wear on the carbon brushes. If the wear on the carbon brushes exceeds one-third of their total length, they should be replaced with new ones promptly.
  2. Desktop centrifuge When the main power indicator light is not illuminated, check whether the indicator light’s fuse and the fuses on the indoor distribution panel have blown. Also, verify that the power cord is making good contact. If a fuse has blown, replace it and ensure the circuit remains unobstructed. Additionally, check the vacuum pump gauge and the oil pressure reading. If the oil pressure is excessively high and the main unit still fails to start, you must inspect all oil lines for blockages, paying particular attention to whether the throttle orifices are clear. If any orifice is blocked, clean it thoroughly to restore proper flow.
  4. If there’s a layer of oxide on the commutator surface, or if the commutator has become uneven due to burning, or if the brushes don’t properly align with the outer edge of the commutator, this can cause the speed to drop. Therefore, the commutator and brushes should be cleaned to ensure good contact.
  5. If there is a short circuit or open circuit in the rotor coil, use a multimeter to check it and then rewind the coil for the medical centrifuge. In the case of benchtop centrifuges, the rotor may rupture, leading to sample leakage and rotor damage during use due to factors such as metal fatigue, overspeeding, excessive stress, chemical corrosion, improper selection, imbalance of the rotor head during operation, and uncontrolled temperature.
  6. If the bearing is damaged or its rotation is obstructed, or if there is insufficient lubrication inside the bearing or excessive dirt buildup causing increased frictional resistance, the motor will fail to reach its rated speed. The bearing should be cleaned or replaced promptly.
  7. Clean the centrifuge chamber and rotor, and wipe dry any condensation inside the chamber. After use, leave the door open until the chamber returns to room temperature. Before centrifuging, ensure that samples are properly balanced. For benchtop centrifuges, remember to disinfect and sterilize after centrifugation. If necessary, precool the rotor. When maintaining and storing rotors and heads, be sure to protect the counting ring at the bottom.
  Shanghai Luxiangyi Centrifuge Instrument Co., Ltd. is stepping up its R&D efforts and working hard to develop new desktop centrifuge products, thereby elevating China’s desktop centrifuge technology to a new level. The latest development... Desktop centrifuge The key technical performance has reached the internationally advanced level of similar products, and both the product’s manufacturing process and its appearance have been significantly improved.