Take the laboratory centrifuge as an example to analyze rotational speed calibration.
Release time:
2021-11-12
Overview:
The calibration of centrifuge rotation speeds has always been highly valued by buyers.
Because many centrifuges are used in various laboratories—including hospital laboratories, research institution laboratories, food and drug testing laboratories, and food hygiene testing laboratories—these institutions all have high requirements for data accuracy, and their rotational speeds do not exceed 30,000 r/min.
Just take Laboratory centrifuge As for the rotational speed, it is a critically important technical parameter. The magnitude of centrifugal force is directly proportional to the rotational speed of the laboratory centrifuge. If the rotational speed fails to meet the specified requirements, the suspension or emulsion cannot be thoroughly separated. During laboratory analysis, using different laboratory centrifuges to separate the same substance may lead to divergent conclusions, seriously affecting the accuracy of the test results.
Therefore, a rotational speed meter is required, based on the metrological performance of laboratory centrifuges. The rotational speed meter must have an accuracy of 0.2% or better, and its expanded uncertainty U(k=2) should meet the calibration requirements for the technical specifications of the laboratory centrifuge being measured (i.e., the expanded uncertainty value should not exceed one-third of the absolute value of the measurement error).
The calibration of centrifuges can generally be divided into two aspects: rotational speed error and rotational speed stability.
1. Calibration of rotational speed error
Before calibration, the measurement equipment should be allowed to fully stabilize under the specified conditions. Set the laboratory centrifuge—equipped with a rotor loaded to its maximum capacity—to its highest rated speed and allow it to stabilize for 5 minutes. Use a tachometer to measure the centrifuge’s rotational speed, taking one reading every minute for a total of 10 readings. Calculate the speed error for each reading according to Formula (1), and take the maximum value among these errors.
Rotational speed error calculation:

2. Rotational Speed Stability Calibration
Set the laboratory centrifuge, loaded to its maximum capacity, to its highest rated speed and allow it to stabilize for 5 minutes. Use a tachometer to measure the centrifuge’s rotational speed, taking one measurement every minute for a total of 10 measurements (alternatively, you may use data recorded from measuring rotational speed errors). Calculate the arithmetic mean of the 10 measured speeds, and then compute the stability of each measurement using Equation (2). Take the maximum value obtained as the final result.
Rotational Speed Stability Calculation:

The above is a brief description of the calibration method for laboratory centrifuge speeds. If you’d like to learn more, please stay tuned to Shanghai Luxiangyi Centrifuge Instrument Co., Ltd.