A method for testing centrifuges using Escherichia coli
Release time:
2014-01-24
Overview:
In microbiology, bacterial testing often takes about a week, and the results are no longer very meaningful for clinical applications or pharmaceutical production—especially in the inspection of pharmaceutical semi-finished products, where such results fail to meet the requirements of drug manufacturing. Although there have been reports of improved methods, these still remain dependent on traditional testing approaches and suffer from poor specificity and sensitivity. Based on the principle of specific antigen-antibody binding and drawing on practical experience, we have successfully tested for E. coli antigens using anti-E. coli antiserum derived from E. coli itself.
1 Experimental Section
1.1 Equipment and Materials
Healthy New Zealand Large-Eared White Domestic Rabbit ( Center for Animal Experimentation, First Military Medical University ) Escherichia coli (44102-15 National Institute for Drug and Biological Products Control, Ministry of Health ) ; Clinically isolated strains of Escherichia coli ( Department of Laboratory Medicine, Nanfang Hospital )。UV-265 UV spectrophotometer; shaker and constant-temperature water bath. Lu Xiangyi Escherichia coli Inspection Centrifuge, shaker, sterilized large test tubes, capillary tubes, slides, syringe tubes. 0.5% Carbonic acid solution, thimerosal solution, saturated ammonium sulfate solution, phosphate buffer, hydrochloric acid solution, sodium hydroxide solution, and others. All materials used in the experiment must be sterilized.
1.2 Methods and Results
1.2.1 Preparation of Escherichia coli antigen – Take 5 ml Carbolic acid saline solution, injected via... 24 h The cultured Escherichia coli slant was washed to remove the bacterial lawn and transferred into a sterile Erlenmeyer flask, where it was... 60 Place in a constant-temperature water bath shaker. 4 h Take out and shake well. Count simultaneously using the turbidimetric method and the petri-dish method, and dilute with carbolic acid saline solution to... 9 × 109 one /ml Put it in again. 60 In a ℃ water bath 1 ~ 1.5 hours Take a small amount for sterile testing, and store the remaining bacterial suspension in the refrigerator.
1.2.2 Preparation of Escherichia coli antiserum: Take a bacterial suspension that has passed the sterility test and proceed according to the table. 1 Shown [ 1 ], administered via intravenous injection into the rabbit ear, and according to the table. 2 Collect serum via the rabbit ear vein. 1 ml, Using Lu Xiangyi Escherichia coli Inspect the centrifuge for Separation, titer determination by the slide method, results: No. 14 The highest daily potency is 1 ∶ 5120 , obtained from rabbit arteries, the first 14 Day blood [ 2 ]Make the largest possible slope, leave it overnight in the refrigerator, collect the serum, and add approximately [amount] of thimerosal solution. 2.5 ml/30 ml Serum.
Table 1 E. coli Antigen Injection Dosage Schedule
Table 1 Schedule of injection time and dosage of E.C. bacillus antigen
Time Dosage (ml) Avoirdupois of rabbit (kg)
One day 0.2 1.8
Two day 0.4 2.0
Four day 0.6 1.9
Six day 2.0 2.0
1.2.3 Antisera Preservation Experiment: The undiluted and diluted antiserum against Escherichia coli were placed in a refrigerator, and their titers were determined at different time points. The results showed that the antiserum could be preserved for at least... 50 d( Table 2) 。
Table 2 Storage Time Experiment
Table 2 Time-tested antiserum kept
Efficiency of
Antiserum firmly extends.
10 days 20 days 30 days 40 days 50 days 60 days
1 ∶ 5120 +++ +++ +++ +++ +++ +++
1 ∶ 2560 +++ +++ +++ +++ +++ +++
1 ∶ 1280 +++ +++ +++ +++ +++ +++
1 ∶ 640 +++ +++ +++ +++ ++ ++
1 ∶ 320 +++ +++ ++ ++ + -
+++、++ Both the slide method and the tube method formed a firm gel. + The slide method gel is firm, while the tube method gel is not firm. - Indicates that neither the slide method nor the tube method causes coagulation.
1.2.4 Preparation of Mixed Antisera: Take Escherichia coli. (44105-15) Mixed antigens and mixed antisera were prepared using strains isolated from clinical specimens and pharmaceutical preparations. The titer of the resulting mixed antisera was: 1 ∶ 4 800。
1.3 Original detection of Escherichia coli
1.3.1 Detect known antigens using Escherichia coli. (44102-15)1 Strain, clinically isolated strain 5 Strain, drug-separated strain 4 Strains and drug-contaminated E. coli (44102-15) Post-separation strain 5 For example, using the slide method with 1 ∶ 320 The titer of the antiserum tested was all positive. ( Both the positive and negative controls are valid. The negative control uses inactivated Staphylococcus and Salmonella. )。
1.3.2 Detecting unknown antigens: The drug under examination and clinical specimens are subjected to... 24 h Culture broth enrichment solution and Lu Xiangyi Centrifugal machine Direct separation (10,000 rpm) The sample solutions were each tested with the pooled antiserum, and the results were consistent with those obtained by the conventional method. ( Table 3) 。
Table 3 Sample measurement results
Table 3 Result of sample checked
Sample (share) This method Traditional method
+ - + -
SP of cell(12) 1 11 1 11
LP of sell(6) 0 6 0 6
SP of self-product (36) 2 34 2 34
LP of self-product (25) 3 22 3 22
Clinic sample(3) 2 1 2 1
SP = Solid preparation ; LP = Liquid preparation
2 Discussion
2.1 Antigen-antibody agglutination is a specific and sensitive method that requires the preparation of antibodies against multiple antigen groups to meet the needs of different antigens. Using mixed antibodies to detect various antigen groups of Escherichia coli can generally satisfy routine laboratory requirements, achieving rapid and accurate results. For bacteria with a single antigen, it suffices to prepare antibodies specifically against that antigen.
2.2 The preparation of mixed antibodies requires careful selection of antigen groups to ensure that the resulting mixed antibodies can detect the specified bacteria and avoid false negatives.
2.3 The method described in the text requires inactivating bacteria to prepare the antigen. In antibody preparation, it is essential to carefully study the injection dose, injection timing, and blood-collection time of the antigen to avoid failure in antigen preparation or suboptimal antibody titers. Moreover, the requirements for bacterial samples are not stringent; whether using centrifugal sedimentation for bacterial collection or enrichment culture, neither approach affects the sensitivity or accuracy of the assay.