Safe Use and Maintenance of the Luxiangyi Centrifuge

Release time:

2014-06-20


Overview:

If you notice any abnormal phenomena in the centrifuge, immediately stop using it and contact us. If the centrifuge displays an error code, please contact our company at 021-68113426.

1. Ensure that the centrifuge is placed on a sturdy, level floor or laboratory workbench, with all four rubber feet making full contact with the surface and distributing weight evenly to prevent vibrations.

2. If the instrument malfunctions, do not disassemble or repair it yourself. If necessary, please contact our company or a qualified professional authorized by the company promptly.

3. When replacing or removing the centrifuge rotor, you must turn off the centrifuge’s power supply to avoid potential hazards.

4. Please ensure that the instrument is properly grounded and has a dedicated ground wire.

5. While the entire machine is running, do not manually open the centrifuge door cover in an emergency. Also, do not move the centrifuge when the rotor is spinning.

6. Before each use, carefully inspect the rotor for cracks or corrosion spots. If any are found, do not use it—please contact our company!

7. Before each use, check whether the screws connecting the rotor to the motor shaft are loose. If they are, tighten them with a screwdriver.

8. For operational safety, maintain at least 30 centimeters of clear space around the centrifuge.

9. If a test tube breaks inside the centrifuge chamber, be sure to thoroughly clean up any fragments before using it again.

10. The centrifuge should be placed in an area that is protected from ultraviolet radiation or direct sunlight to prevent discoloration and paint chipping on its exterior.

11. If you notice any abnormal phenomena in the centrifuge, immediately stop using it and contact us. If the centrifuge displays an error code, please contact our company at 021-68113426.

12. It is strictly prohibited to use rotors at speeds exceeding the specified limits. Users shall be solely responsible for any rotor-related accidents or losses resulting from failure to set the rotor number according to the instructions or from operating the rotor at speeds beyond its rated capacity.

13. When the machine is to be left unused for an extended period, the rotor should be removed to prevent the spindle from rusting.

14. After the rotor is removed from the centrifuge chamber, promptly wipe it dry with a neutral detergent to prevent chemical corrosion. Store it in a dry, well-ventilated area. Do not use non-neutral cleaning agents to clean the rotor. Apply a small amount of lubricant to the central hole of the rotor for protection.

15. The rotor installation and removal procedures should be performed gently. When removing the rotor, lift it vertically upward; otherwise, the motor shaft may easily become bent.

16. If the centrifuge will not be used for an extended period, clean the inner cavity, dry it thoroughly, and apply a small amount of grease to the motor shaft for protection.

17. Do not place any items on the door cover to prevent damage to those items if the door cover opens automatically.

18. If a centrifuge tube breaks during operation, causing significant vibration, the machine should be stopped immediately.

19. After each use, remove any accumulated water at the bottom of the centrifuge chamber that has resulted from adding an excessive amount of sample.