Do you know the correct way to properly open a low-speed centrifuge?
Release time:
2018-11-06
Overview:
Low-speed centrifuge This equipment exhibits strong adaptability to various materials. By selecting the appropriate filter medium, it can effectively separate fine particles as small as millimeters in size and is also suitable for dewatering bulk items. Cleaning of the items can be carried out via a water-washing pipe. The manually operated top-discharge design features a simple structure, easy installation, low cost, and straightforward operation, while preserving the crystal shape of the product. The machine employs an elastic support structure that helps reduce vibrations caused by uneven loading, ensuring smooth operation. The entire mechanical assembly is housed within a sealed enclosure, enabling tight sealing and preventing contamination of the materials.
Traditional centrifuges feature an open-type drive belt, with only a protective cover at the motor end—primarily for safety reasons to enhance operational safety. However, this design does not control the dust generated by friction, making it unsuitable for use in cleanroom environments. Therefore, it is essential to seal off the drive belt’s operating area to prevent the escape of frictional dust. These centrifuges are suitable for washing and separating solid and liquid phases in suspensions containing solid particles, as well as for solid-liquid separation in media containing crystalline solids. Low-speed centrifugation technology represents a new approach developed based on the behavior of particles under practical centrifugal conditions. Particles of varying densities, sizes, or shapes settle differently depending on the centrifugal speed; thus, even a roughly spherical, non-homogeneous mixture can be separated using centrifugation. Benchtop low-speed centrifuges are designed for further research in biochemistry and for the large-scale separation of various substances—for example, collecting cells, isolating plasma, purifying preparative samples to separate DNA and protein molecules, and also enabling the isolation of viruses, large-scale E. coli cultures, cellular components, and nucleoprotein particles.
The rotor of the low-speed centrifuge should be securely fastened in the correct position, and the fixing screws must be tightened. Check whether the rotor and other accessories have cracks or signs of corrosion, and ensure that the ground wire is making proper contact. Always use a dedicated power outlet to guarantee stable voltage. If the user’s voltage is unstable, it is essential to connect a voltage stabilizer to prevent damage to the centrifuge. The benchtop low-speed centrifuge must be placed on a sturdy, stable, and level work surface, with sufficient space left around the unit to ensure good ventilation. After each use, open the lid, wipe off any condensation, and allow the unit to air-dry naturally. Before and after centrifugation, the rotor must be carefully lowered or lifted vertically to avoid any impact on the spindle or rotor itself. After use, the centrifuge rotor, basket, sleeves, and other components should be thoroughly dried and stored separately.
During the centrifugation process with a low-speed centrifuge, if a glass tube breaks, all debris—including fragments from the centrifuge chamber and sleeves—must be thoroughly removed. Otherwise, the centrifuge could be damaged. Regularly use a vacuum cleaner to remove dust from the heat sinks located at the rear of the centrifuge. If you notice any corrosion or cracks on the bricks of the low-speed centrifuge, they should be replaced immediately. Additionally, it is essential to periodically maintain the rotor, basket, and sleeves with specialized polishing oil to prevent corrosion, and to lubricate components such as the shaft and basket ears with appropriate lubricant.
Proper use of a low-speed centrifuge can help prevent many accidents, such as:
Load the sample and find balance.
During the design and manufacturing of centrifuges, any imbalance caused by machining errors in the rotor has been corrected through dynamic balancing tests. However, every centrifuge has a specified allowable amount of sample imbalance. It is beneficial for the lifespan of the centrifuge to achieve as good a balance as possible before starting the centrifugation process.
Clean the accumulated water in the centrifuge chamber.
When the centrifuge is running, refrigeration causes moisture in the air to condense into frost inside the centrifuge chamber. After the machine is stopped, this frost melts into water. Most low-speed, high-capacity centrifuges manufactured abroad do not have drain holes, so water gradually accumulates in the centrifuge chamber. In such cases, users should remove the rotor themselves and clear out the accumulated water. When reinstalling the rotor, be sure to secure it properly to prevent accidents.
Aluminum alloys are not susceptible to corrosion.
Centrifuge rotors are typically made of aluminum alloy. Once corroded, their strength diminishes, making them prone to accidents. Aluminum alloys are susceptible to liquid corrosion; after cleaning, they should be dried thoroughly using a hairdryer or by inverting them for a period of time—only use them once you’ve confirmed they’re completely dry. Some blood banks’ imported centrifuges lack the six plastic holders that should normally be included in the six aluminum cups. These plastic holders are meant to hold blood bags, preventing them from being placed directly into the aluminum cups. Also, pay close attention to whether the test tubes have any cracks. If a test tube—especially one that’s been used repeatedly—shows even the slightest crack, never use it under any circumstances. Otherwise, during operation, a cracked test tube could cause an imbalance during rotor rotation, potentially leading to serious accidents such as shaft breakage.
Shaft failure accident
If the rotor is not properly installed or if the sample loading is severely unbalanced, turning on the centrifuge can be extremely dangerous. Although modern imported centrifuges are equipped with imbalance protection—meaning they will automatically cut power and stop the rotor once the imbalance exceeds a certain threshold—it’s already too late by then, and a serious accident could occur. One such severe accident is shaft failure: when the shaft suddenly breaks during high-speed rotation, the centrifuge rotor loses its support and starts spinning wildly inside the chamber, potentially causing the entire centrifuge to rotate by as much as 270 degrees. At this point, if there happens to be personnel or equipment nearby, it could lead to injuries or even the destruction of valuable equipment.
When selecting a centrifuge, users must determine the model and type of low-speed centrifuge based on the physicochemical properties of the separation medium (including medium type, temperature, concentration, viscosity, slippage of solid particles, filter cake hardness, requirements for solid-phase particles, and allowable residual liquid phase in the solid-phase filter cake), as well as the corrosion- and explosion-proof requirements applicable during the separation process and in the specific production environment, in order to meet both the intended purpose and safety standards.