Find the cause from the results—centrifuge design prioritizes safety.

Release time:

2020-09-11


Overview:

The design, selection, and installation of centrifuges shall comply with pharmaceutical manufacturing requirements, be easy to clean, disinfect, or sterilize, facilitate production operations as well as maintenance and repair, and effectively prevent errors or reduce contamination. Surfaces of centrifuges that come into direct contact with pharmaceutical products shall be smooth, flat, free of dead corners, easy to clean or disinfect, corrosion-resistant, and neither chemically react with nor adsorb the pharmaceutical products.

   Centrifuge The design, selection, and installation of equipment shall meet pharmaceutical production requirements, be easy to clean, disinfect, or sterilize, facilitate production operations as well as maintenance and repair, and effectively prevent errors or reduce contamination. The surfaces of centrifuges that come into direct contact with pharmaceuticals shall be smooth, flat, free of dead corners, easy to clean or disinfect, corrosion-resistant, and neither chemically react with nor adsorb the pharmaceutical products.

  The lubricants, coolants, and other substances used in centrifuges must not contaminate the pharmaceutical products or containers. The design and installation of pipelines should avoid dead ends and blind pipes. Personnel entering cleanrooms (or areas) must not wear makeup or jewelry, nor may they directly touch pharmaceutical products with bare hands. Instruments, meters, measuring tools, weighing devices, and other equipment used for production and inspection must meet the applicable scope and precision requirements for production and inspection, bear clear certification marks indicating their compliance, and undergo regular calibration. Centrifuges used in production must have clearly visible status indicators and be regularly maintained, serviced, and verified. Operations related to the installation, maintenance, and servicing of centrifuges must not compromise product quality. For instruments and meters with special requirements, they should be housed in dedicated instrument rooms equipped with facilities that protect against static electricity, vibration, humidity, or other external factors.

  To improve the utilization rate and extend the service life of laboratory instruments, ensure their safe operation, and prevent accidents, this article analyzes the factors contributing to accidents involving laboratory centrifuges.

  Clean the accumulated water in the centrifuge chamber.

  For some centrifuges equipped with refrigeration functions, moisture in the air can cause frost to form inside the centrifuge chamber during operation. After the machine is turned off, this frost melts into water. Most low-speed, high-capacity centrifuges do not have drain holes, so water tends to accumulate more and more in the centrifuge chamber. In such cases, users should remove the rotor themselves and clear out the accumulated water. When reinstalling the rotor, be sure to secure it properly to prevent accidents.

  Aluminum alloys are not susceptible to corrosion.

  Centrifuge rotors are typically made of aluminum alloy. Once corroded, their strength diminishes, making them prone to accidents. Aluminum alloys are susceptible to liquid corrosion; after cleaning, they should be dried thoroughly using a hairdryer or by turning them upside down for a period of time—only after confirming that they’re completely dry should they be put into use. Some blood banks’ imported centrifuges lack the six plastic holders that should normally be included in the six aluminum cups. These plastic holders are meant to hold blood bags, preventing them from being placed directly inside the aluminum cups. Also, pay close attention to whether the test tubes have any cracks. If a test tube—especially one that has been used repeatedly—shows even the slightest crack, it must never be used. Otherwise, during operation, a cracked test tube could cause an imbalance during rotor rotation, potentially leading to serious accidents such as shaft breakage.

  Shaft failure accident

  If the rotor is not properly installed or if the sample loading is severely unbalanced, the centrifuge will be started. Although modern imported centrifuges are equipped with imbalance protection—meaning that when the imbalance exceeds a certain threshold, the machine will automatically shut down—the damage is already done in the scenarios described above, and a serious accident may occur. One such severe accident is shaft failure: due to the sudden breakage of the shaft during high-speed rotation, the centrifuge rotor loses its support and starts spinning wildly, colliding randomly within the centrifuge chamber. At this point, if there are personnel or equipment nearby, it could lead to injuries and destruction of equipment.

  The above is a brief summary of centrifuge operation. If you’d like to learn more about centrifuges, please stay tuned for updates from Shanghai Luxiangyi Centrifuge Instrument Co., Ltd.