How to Choose a Rotor When Purchasing a Centrifuge

Release time:

2017-03-27


Overview:

  Buy Centrifuge Before using a centrifuge, it’s essential to first familiarize yourself with its component parts and related operating procedures. This will help extend the centrifuge’s service life. The rotor is one of the centrifuge’s critical components and is closely linked to safety requirements in the production process. Here, Shanghai Luxiangyi Company would like to remind all users: Don’t be tempted by low prices and purchase substandard centrifuges. When buying a centrifuge, be sure to select the rotor thickness and other operational parameters according to the manufacturer’s specifications.
  Safety requirements for centrifuge use:
  1. The selection must be correct.
  2. Use the centrifuge within the specified range. Do not use centrifuges intended for general purposes in flammable, explosive, or corrosive environments.
  3. Operate according to the prescribed operating procedures; do not engage in abnormal braking, overload operation, or overspeed operation.
  4. Faults should be promptly addressed; operation with faults is prohibited. Centrifuge Promptly scrap centrifuges that have become obsolete or failed.
  5. Install, use, and maintain the centrifuge according to the instructions provided in the product manual. Do not modify the centrifuge’s components on your own, nor increase its rotational speed without authorization.
  6. Develop and strictly enforce relevant rules and regulations for the installation, operation, and maintenance of centrifuges to ensure safe use.
  How to select a rotor when purchasing a centrifuge—this can be calculated using the following formula:
  Once the capacity is determined, the most important parameter—not the rotational speed but the centrifugal force—becomes the key factor. The magnitude of the centrifugal force directly determines the effectiveness of separation. Although rotors of the same capacity from different manufacturers may achieve the same rotational speed, their centrifugal forces are not necessarily identical. Some manufacturers, in order to reduce production costs and minimize noise, design rotors with particularly small angles (a smaller angle results in a smaller centrifugal radius). As a result, at the same rotational speed, such rotors generate a lower centrifugal force compared to those with larger angles, and it also becomes less convenient to collect the separated material.
  Attachment: Centrifugal Force Calculation Formula:
  RCF = 11.2 × R × (r/min / 1000)²
  The above formula is the calculation formula for centrifugal force (RCF), where R represents the radius of the rotor classification, and r/min represents the rotational speed.
  If you would like to learn more Centrifuge For related information, please visit the official website of Shanghai Luxiangyi. Our editor will update relevant information regularly.