A Glimpse into the Tip of the Iceberg: Common Faults in Medical Centrifuges
Release time:
2015-10-08
Overview:
Medical centrifuge
Epigraph: Typical malfunctions in medical centrifuges are relatively common. Today, I’d like to offer a brief analysis on this topic. Please note that these are merely personal insights and may not be entirely comprehensive—please bear with any shortcomings. We’d also welcome your calls to Shanghai Luxiangyi Centrifuge Instrument Co., Ltd. or visits to our company website to join us in discussing this further. Together, let’s strive to advance the development of medical centrifuges.
In medical laboratory testing, medical centrifuges are commonly used instruments for separating serum, plasma, precipitating proteins, or performing urine sediment analysis. By utilizing a medical centrifuge, suspended particles in a mixed liquid can be rapidly precipitated, thereby enabling the separation of various components with different specific gravities.
In laboratory testing, medical centrifuges are frequently started and stopped, with usage concentrated over relatively short periods. As a result, the motors and mechanical components experience significant wear and tear. If maintenance and repairs are not performed regularly and on time, the centrifuge is highly likely to malfunction—such as failing to start or running at a reduced speed—directly disrupting the normal conduct of testing work. Therefore, it is crucial to master methods for promptly and efficiently troubleshooting various centrifuge malfunctions, ensuring that the centrifuge remains in perfect working order at all times.
Troubleshooting: If a medical centrifuge stops rotating or rotates slowly, take the following steps for troubleshooting.
Cause Analysis: After the medical centrifuge is powered on, gradually increase the speed control knob clockwise, one gear at a time, to steadily accelerate the motor’s rotational speed until it reaches the desired operating speed. When the centrifuge malfunctions—such as failing to start or running at a slow speed—the first step is to rule out any faults in the power supply section. Aside from issues like broken power cords, poor contact between plugs and sockets, damaged bearings, short circuits or open circuits in the stator or rotor windings, or burnout, the primary cause is typically an open circuit in the brushes—that is, inadequate contact and poor connection between the commutator on the rotor and the brushes.
Because Medical centrifuge The motors used in the above applications are generally series-wound motors. They consist of two main parts: the stator and the rotor. The stator comprises a centralized, fixed magnetic pole made up of an iron core and magnetic field coils. The rotor consists of an iron core, rotor coils, and a commutator. The brushes serve as crucial components that connect the stator coils to the rotor coils via the commutator. The operating principle of the series-wound motor's circuit is as follows: current flows from the power source through the stator’s magnetic field coils, then via the brushes and through the commutator into the rotor coils, and finally returns to the power source through the brushes and the stator’s magnetic field coils, thus forming a closed circuit. The magnetic field coils and rotor coils are connected in series through the commutator and brushes. Since the current is the same at every point in the circuit, the motor can operate normally at the desired speed only if the series circuit remains unobstructed and free of any blockages.
As can be seen from this, the operating condition of the carbon brushes directly affects the normal operation of the motor. However, due to factors such as wear and tear, consumption, or spring failure during the contact between the brushes and the commutator, the brushes are prone to open circuit and lose contact with the commutator, which is a major cause of centrifuge shutdowns or slow rotation speeds.
Maintenance Steps: First, clean the brush holder and inspect the brushes for wear. When the length of a worn brush falls below 10 mm, it should be replaced with a brush of the same specification. To replace the brush, unscrew the brush holder cover and remove the old brush. If the spring on the old brush is intact, you can detach it and reattach it to the new brush. If the spring is失效 or damaged, replace it at the same time. Then, align the brush in the correct direction and reinsert it into the holder as it was originally, ensuring that the brush drops freely into the holder with appropriate tightness. Finally, screw the holder cover back tightly. Next, remove the rotating disc, set the motor speed to its maximum, and let the motor run unloaded. At this point, the motor speed will be very high; after a few minutes, the end face of the brush will align perfectly with the outer edge of the commutator. At this stage, you can stop the motor, disassemble it, and use fine sandpaper to remove any oxides from the commutator surface. Then, use a brush or dry cloth to thoroughly remove all carbon dust particles from the commutator. Finally, reassemble the machine and restore it to its original condition.
Another simple method is to find a cylinder roughly the same size as the commutator, wrap it with fine sandpaper—with the abrasive grains facing outward—and first grind the end face of the brush into an arc that perfectly matches the outer edge of the commutator. Then, reassemble the brush. To ensure the centrifuge operates smoothly, it’s crucial to perform thorough preventive maintenance regularly—every six months or once a year, check the wear condition of both the commutator and the brushes. Promptly remove any carbon dust or debris from the commutator surface to extend the service life of the brushes. For damaged components such as brushes and springs, replace them promptly to keep the centrifuge in optimal working condition. Anatomical sites: (4) highly irregular tumors; (5) tumors located adjacent to normal tissue structures that are radiosensitive; (6) small-volume tumors requiring high-dose radiation. Conformal radiotherapy is not suitable for tumors that are either extremely sensitive or extremely resistant to radiation therapy.
Shanghai Luxiangyi Centrifuge Instrument Co., Ltd. is a Chinese company that has obtained FDA certification for its centrifuges and specializes in the manufacture of... Medical centrifuge Laboratory centrifuges, mini centrifuges, cell centrifuges, high-speed centrifuges, Xiangyi centrifuges, graphene centrifuges, and rotor accessories—backed by years of market sales experience and a wealth of in-depth technical expertise—we warmly welcome both new and existing customers to call for inquiries. We’re committed to providing you with the utmost dedication and service!
In medical laboratory testing, medical centrifuges are commonly used instruments for separating serum, plasma, precipitating proteins, or performing urine sediment analysis. By utilizing a medical centrifuge, suspended particles in a mixed liquid can be rapidly precipitated, thereby enabling the separation of various components with different specific gravities.
In laboratory testing, medical centrifuges are frequently started and stopped, with usage concentrated over relatively short periods. As a result, the motors and mechanical components experience significant wear and tear. If maintenance and repairs are not performed regularly and on time, the centrifuge is highly likely to malfunction—such as failing to start or running at a reduced speed—directly disrupting the normal conduct of testing work. Therefore, it is crucial to master methods for promptly and efficiently troubleshooting various centrifuge malfunctions, ensuring that the centrifuge remains in perfect working order at all times.
Troubleshooting: If a medical centrifuge stops rotating or rotates slowly, take the following steps for troubleshooting.
Cause Analysis: After the medical centrifuge is powered on, gradually increase the speed control knob clockwise, one gear at a time, to steadily accelerate the motor’s rotational speed until it reaches the desired operating speed. When the centrifuge malfunctions—such as failing to start or running at a slow speed—the first step is to rule out any faults in the power supply section. Aside from issues like broken power cords, poor contact between plugs and sockets, damaged bearings, short circuits or open circuits in the stator or rotor windings, or burnout, the primary cause is typically an open circuit in the brushes—that is, inadequate contact and poor connection between the commutator on the rotor and the brushes.
Because Medical centrifuge The motors used in the above applications are generally series-wound motors. They consist of two main parts: the stator and the rotor. The stator comprises a centralized, fixed magnetic pole made up of an iron core and magnetic field coils. The rotor consists of an iron core, rotor coils, and a commutator. The brushes serve as crucial components that connect the stator coils to the rotor coils via the commutator. The operating principle of the series-wound motor's circuit is as follows: current flows from the power source through the stator’s magnetic field coils, then via the brushes and through the commutator into the rotor coils, and finally returns to the power source through the brushes and the stator’s magnetic field coils, thus forming a closed circuit. The magnetic field coils and rotor coils are connected in series through the commutator and brushes. Since the current is the same at every point in the circuit, the motor can operate normally at the desired speed only if the series circuit remains unobstructed and free of any blockages.
As can be seen from this, the operating condition of the carbon brushes directly affects the normal operation of the motor. However, due to factors such as wear and tear, consumption, or spring failure during the contact between the brushes and the commutator, the brushes are prone to open circuit and lose contact with the commutator, which is a major cause of centrifuge shutdowns or slow rotation speeds.
Maintenance Steps: First, clean the brush holder and inspect the brushes for wear. When the length of a worn brush falls below 10 mm, it should be replaced with a brush of the same specification. To replace the brush, unscrew the brush holder cover and remove the old brush. If the spring on the old brush is intact, you can detach it and reattach it to the new brush. If the spring is失效 or damaged, replace it at the same time. Then, align the brush in the correct direction and reinsert it into the holder as it was originally, ensuring that the brush drops freely into the holder with appropriate tightness. Finally, screw the holder cover back tightly. Next, remove the rotating disc, set the motor speed to its maximum, and let the motor run unloaded. At this point, the motor speed will be very high; after a few minutes, the end face of the brush will align perfectly with the outer edge of the commutator. At this stage, you can stop the motor, disassemble it, and use fine sandpaper to remove any oxides from the commutator surface. Then, use a brush or dry cloth to thoroughly remove all carbon dust particles from the commutator. Finally, reassemble the machine and restore it to its original condition.
Another simple method is to find a cylinder roughly the same size as the commutator, wrap it with fine sandpaper—with the abrasive grains facing outward—and first grind the end face of the brush into an arc that perfectly matches the outer edge of the commutator. Then, reassemble the brush. To ensure the centrifuge operates smoothly, it’s crucial to perform thorough preventive maintenance regularly—every six months or once a year, check the wear condition of both the commutator and the brushes. Promptly remove any carbon dust or debris from the commutator surface to extend the service life of the brushes. For damaged components such as brushes and springs, replace them promptly to keep the centrifuge in optimal working condition. Anatomical sites: (4) highly irregular tumors; (5) tumors located adjacent to normal tissue structures that are radiosensitive; (6) small-volume tumors requiring high-dose radiation. Conformal radiotherapy is not suitable for tumors that are either extremely sensitive or extremely resistant to radiation therapy.
Shanghai Luxiangyi Centrifuge Instrument Co., Ltd. is a Chinese company that has obtained FDA certification for its centrifuges and specializes in the manufacture of... Medical centrifuge Laboratory centrifuges, mini centrifuges, cell centrifuges, high-speed centrifuges, Xiangyi centrifuges, graphene centrifuges, and rotor accessories—backed by years of market sales experience and a wealth of in-depth technical expertise—we warmly welcome both new and existing customers to call for inquiries. We’re committed to providing you with the utmost dedication and service!