An Analysis of the Importance of the Centrifuge Rotor as a Centrifugation Medium in High-Speed Refrigerated Centrifuges
Release time:
2020-04-28
Overview:
High-speed refrigerated centrifuge During the R&D phase, rigorous measurements were conducted on the temperature differences between the interior of the centrifugal chamber and the sample under various conditions. Based on these test results, a specialized temperature-compensation model was designed and integrated into the central software system of the centrifuge, along with the sensor.
When substances precipitate in a medium, they are accompanied by diffusion. Diffusion is unconditional and absolute; the smaller the particles of a substance, the greater the range of diffusion. In contrast, sedimentation is conditional and requires the action of external forces. The greater the mass of an object, the faster its sedimentation rate. Since some particles are so small that their sedimentation process is difficult to observe, high-speed refrigerated centrifuges have been developed.
The high-performance refrigeration system of the high-speed refrigerated centrifuge ensures that the rotor’s temperature remains within the required range even during high-speed operation. An excellent refrigeration and temperature-control system is also a standard feature of OHAUS refrigerated centrifuges.
Centrifugation methods, centrifugation media, and density gradients can all affect the separation efficiency of a centrifuge. In addition to these factors, the temperature and pH of the centrifugation medium, the centrifugal speed and centrifugal force, and the centrifugation time are also critical to achieving optimal separation results. For biological samples, in order to prevent aggregation, denaturation, and inactivation of the substances to be separated, it is essential not only to pay careful attention to the selection of the centrifugation medium but also to carefully control other centrifugation conditions, such as temperature and the pH of the medium solution.
Centrifugal medium temperature control
Centrifugal temperature is generally controlled at around 4℃.
For certain biological samples with good thermal stability, the centrifugation process can be carried out at room temperature.
Note: However, when operating at excessive speeds or during high-speed centrifugation, the rotor’s rapid rotation generates heat, leading to a rise in temperature and potential sample loss. Therefore, it is essential to employ a cooling system to maintain the temperature within a stable range.
Centrifugal Medium pH Control
The pH of the centrifugation medium solution should fall within the pH range that allows the sample to remain stable. If necessary, a buffer solution can be used. Additionally, excessively acidic or alkaline solutions may cause corrosion of the rotor and other components of the centrifuge; therefore, direct contact with these components should be avoided as much as possible, and any such contact should be promptly followed by thorough cleaning.
High-speed refrigerated centrifuges feature a powerful cooling system and temperature compensation capabilities, enabling accurate, reliable, and high-quality results.
In addition, the centrifuge rotor is also an important influencing factor.
The rotor is an important component of a centrifuge. Rotor types include angle rotors, horizontal rotors, and continuous-flow rotors.
During the design and manufacturing of centrifuges, any imbalance caused by machining errors in the rotor is corrected through dynamic balancing tests. However, every centrifuge has a permissible limit for sample loading imbalance. While the centrifuge can still operate within this larger tolerance range, the resulting unbalanced forces will suddenly impact the bearings and support structures at a frequency of n times per minute, thereby causing damage to the centrifuge. Therefore, for those expensive centrifuges, users should strive to achieve optimal balance before starting the centrifugation process—this practice will significantly extend the lifespan of the centrifuge.
Angular rotor centrifuges have a small radius of rotation and high rotational speed, resulting in relatively lower capacity compared to horizontal rotors. They are well-suited for separating and extracting fine particles. Horizontal rotors, on the other hand, are ideal for separating larger particles; they operate at lower speeds than angular rotors but offer greater capacity. Continuous-flow rotors enable quantitative and continuous production, but they are not suitable when the solid content is high. This is because loading and unloading continuous-flow rotors and retrieving the sediment can be cumbersome, time-consuming, and require operators with extensive experience and considerable arm strength. If the solid content exceeds 10%, it is not recommended to choose this type of rotor.
After purchasing a high-speed refrigerated centrifuge, the manufacturer will typically come to your facility for installation and commissioning, as well as provide training for the operators. At this stage, some pharmaceutical companies may not yet have actual materials requiring centrifugation—instead, they’ll just perform a simple setup and basic debugging of the equipment. Once it’s time to start centrifuging a batch, first carefully read the instruction manual to reinforce your understanding. Then, select the appropriate centrifugal force and corresponding rotor speed, and determine the correct centrifugation duration. The choice of centrifugation time should align with the pharmaceutical company’s specific process requirements: too long a duration reduces efficiency, while too short a duration can compromise the separation effect. Therefore, it’s crucial to choose a centrifugation time that’s well-suited to the characteristics of the actual material being processed.
For high-speed refrigerated centrifuges, pharmaceutical manufacturers typically set the temperature at 4°C unless otherwise specified. When setting the acceleration and deceleration rates, it’s generally recommended to choose acceleration level 4 and deceleration level 5 during centrifuge installation and adjustment. However, some specialized pharmaceutical applications may require adjusting the deceleration time to the slowest setting, while others may even demand a specially designed deceleration curve within the program. All such requirements should be clearly stated at the time of purchase.