The cream separator uses the Gerber method to determine the cream fat content.

Release time:

2014-09-10


Overview:

Place the cream meter into a water bath at 65–70°C for 5 minutes. Then, use a cream centrifuge to spin at 1000 r/min for 5 minutes, and subsequently place it back into the 65–70°C water bath.

  Add 10 ml of sulfuric acid (90%) into the Gerber cream separator. Using a special 1 ml milk pipette, draw up a milk sample to the marked level and add it into the cream separator. Then, add 1 ml of isoamyl alcohol, tightly seal the container with a rubber stopper, and shake vigorously to dissolve the milk curds.

Place the fat meter into a water bath maintained at 65–70°C for 5 minutes. Then, using a fat centrifuge, spin it at 1000 r/min for 5 minutes, and subsequently place it back into the 65–70°C water bath. After 5 minutes, remove the fat meter, dry it thoroughly, and read the measurement from the concave bottom edge of the scale mark on the fat column; this reading represents the percentage of fat.