Analysis of the Causes of Test Tube Breakage in Low-Temperature Centrifuges

Release time:

2016-02-16


Overview:

Low-temperature centrifuge
   Low-temperature centrifuge Also known as a refrigerated centrifuge, this model can be categorized into two types based on rotational speed: low-temperature high-speed centrifuges and low-temperature low-speed centrifuges. This centrifuge features microcomputer control, a touch-panel interface, and a large-screen LCD display powered by a frequency-conversion motor. It is equipped with a genuine Taikang compressor unit that ensures smooth operation, along with an environmentally friendly refrigerant suitable for medium temperatures. It offers 10 different acceleration and deceleration settings, allowing users to adjust parameters at any time without needing to stop the machine. Additionally, it supports short-duration centrifugation and is compatible with a variety of rotor capacities and adapters.
  A low-temperature centrifuge is a biochemical instrument that uses the centrifugal force generated during rotation to rapidly precipitate suspended particles in a solution. It is most commonly used in fields such as medicine and public health prevention and control.
  However, low-temperature centrifuges often fail during operation due to broken test tubes, sometimes resulting in the loss of valuable samples and even triggering medical disputes in clinical settings. To address this issue, Shanghai Luxiangyi Centrifuge Instrument Co., Ltd. has summarized its understanding of the causes of test-tube breakage as well as its experience in troubleshooting these issues, aiming to help operators and maintenance personnel better use and maintain this equipment.
  According to reports, among the 217 instances of malfunctions in centrifuges imported into our country, 20% were caused by improper operation, and 10% were related to human factors. This clearly shows that even with advanced medical equipment, human factors cannot be overlooked.
  Practice has shown that, Low-temperature centrifuge A test tube break is more often related to human factors. Common issues include:
  1. For samples placed in symmetrically arranged tubes with unbalanced loads, the weights must be adjusted using a balance, including the test tubes themselves, to achieve equilibrium. If only one test tube is used to hold the sample, the corresponding symmetrical tube must be filled with water to maintain balance; otherwise, during centrifugation, the low-temperature centrifuge may vibrate excessively, potentially causing the tubes to rupture. It’s easy to overlook that centrifuge cups in horizontal rotors can become unbalanced after prolonged use or replacement. Therefore, the centrifuge cups themselves must also be balanced. For example, a centrifuge that has been in use for over 10 years might experience frequent tube ruptures that are difficult to troubleshoot. Later investigation revealed that four of the centrifuge cups had become significantly unbalanced. After replacing these cups and re-balancing them, the problem was finally resolved.
  2. Accelerating too quickly can easily cause the centrifuge tube to rupture due to centrifugal force.
  3. If the low-temperature centrifuge’s rotation speed is too high and exceeds the centrifugal force that the centrifuge tubes can withstand, the tubes may rupture.
  Shanghai Luxiangyi Centrifuge Instrument Co., Ltd. kindly reminds you to choose models with lower noise levels whenever possible. Low-temperature centrifuge This helps maintain a comfortable experimental environment. You also need to be careful when selecting accessories—some experiments require special centrifuge tubes (for centrifuging toxic samples or samples that demand ultra-high-speed centrifugation). Such centrifuge tubes must be used with matching tube sleeves to ensure greater safety. Additionally, there are other specialized sample containers (such as irregular-shaped sample bottles and blood bags). All these details and accessories should be carefully considered when choosing a centrifuge; otherwise, normal operation will not be possible.