Factors Contributing to Centrifuge Tube Rupture

Release time:

2015-05-08


Overview:

Samples placed in the symmetrically arranged holes of the centrifuge must be balanced in weight, together with their corresponding test tubes, using a balance.

1. Load imbalance

Samples placed in symmetrical holes of the centrifuge must be balanced in weight using a balance, along with their corresponding test tubes. If only one test tube is used to hold the sample, the symmetrical tube must be filled with water to achieve balance; otherwise, during centrifugation, the instrument will vibrate, potentially causing the test tube to break.

What’s often overlooked is that centrifuge tubes used for horizontal rotor applications can become unbalanced over time or after being replaced. Therefore, the centrifuge tubes themselves must also be balanced. For example, a centrifuge that had been in use for more than 10 years experienced frequent tube breakage, a problem that proved difficult to troubleshoot. It was later discovered that four of the centrifuge tubes had become significantly heavier on one side. After re-balancing the weights of the centrifuge tubes, the malfunction finally disappeared.

2. Excessive speed

If the centrifuge’s rotation speed is increased dramatically, the instrument will vibrate violently and damage the test tubes. Additionally, if the centrifuge’s rotation speed exceeds the maximum centrifugal force that the centrifuge tubes can withstand, the tubes may rupture. Therefore, you should select and replace centrifuge tubes that are capable of withstanding higher rotation speeds.