Precautions for Using a Decapping Centrifuge
Release time:
2014-12-02
Overview:
Precautions
1. After the centrifuge has completed its run, remove the rotor and place it upside down on the lab bench. Wipe dry any remaining water inside the chamber, and leave the centrifuge lid open.
2. Place the rotor lid either on the centrifuge’s platform or on the lab bench. Under no circumstances should you leave it floating loosely on the rotor without tightening it securely, because if the centrifuge is accidentally started, the rotor lid could fly off and cause an accident!
3. After tightening the rotor cover, be sure to use your fingers to check that there is no gap between the rotor and the rotor cover. If there is a gap, loosen and retighten the cover until you confirm that there is no gap before starting the centrifuge.
4. During the centrifugation process, operators must not leave the centrifuge room. In the event of an abnormal situation, operators must not turn off the power supply (POWER); instead, they should press STOP.
5. Do not use substandard or counterfeit centrifuge tubes; do not use centrifuge tubes that are aged, deformed, cracked, or fail to meet quality standards.
6. During holidays and in the evening, the last staff member to use the centrifuge must conduct a routine safety inspection before leaving.
7. If the instrument malfunctions or components are damaged during use, please contact us—the manufacturer of Luxiangyi centrifuges—immediately. Under no circumstances should you attempt to disassemble the instrument yourself.