Measures to Prevent Damage to Centrifuges at the Beijing Blood Center

Release time:

2013-11-29


Overview:

When placing the centrifuge cup into the centrifuge, to maintain the shape of the blood bag, position the blood bag parallel to the centrifuge’s rotor arm.
Measures to prevent centrifugal damage:
  1. Use the dedicated centrifugal assist cup.
  When performing centrifugation, be sure to use the dedicated centrifuge assist cup! The specialized centrifuge assist cup is designed to minimize the "dead space" within the centrifuge tube; its bottom is rounded to perfectly match the shape of the blood bag. Centrifuge tubes with flat bottoms cannot effectively eliminate dead space, thus significantly increasing the likelihood of tube breakage during centrifugation.
  2. The label faces outward.
  The front side of the blood bag (the side with the label attached) is more likely to suffer centrifugal damage than the back side (the side without the label). This is because, during centrifugation, as the sheet material is stretched, the back side is uniformly elongated throughout, whereas on the front side, only the lower portion—where the label is absent—is stretched. When using a dedicated centrifuge assist cup, the dead space on the outer side of the cup is almost completely eliminated. If the labeled side faces outward toward the exterior of the centrifuge cup, the risk of damage is further reduced.
  3. The Y-tube (tee) and hemostat should be placed in the upper part of the cup.
  When hard objects such as Y-tubes and hemostatic clamps are placed at the bottom of the centrifuge cup, they not only create dead space but also exert pressure on the blood bag during centrifugation, increasing the likelihood of damage. Therefore, it’s best to place Y-tubes and hemostatic clamps as high up as possible in the centrifuge cup. Moreover, if all catheters are positioned in the upper part of the centrifuge cup—not just the Y-tube—the overall effect will be even better.
  4. The final step is “tapping.”
  To minimize dead space, the final step should involve tapping. Place the centrifuge tube containing the blood bag with its bottom facing downward and tap it several times against a hard surface such as a laboratory bench. This will ensure that the bottom of the blood bag makes full contact with the bottom of the centrifuge tube, thereby reducing dead space. The bottom of the specialized centrifuge tube features a fan-shaped arch designed to enhance its strength, so you don’t need to worry about the tube cracking from tapping.
  5. Place the blood bag and the centrifuge arm parallel to each other.
  When placing the centrifuge cup into the centrifuge, to maintain the shape of the blood bag, position the blood bag parallel to the centrifuge’s rotor arm.
6. Be sure to select an appropriate centrifugal force—do not use one that is too high.