The Correct Method for Using a Laboratory Centrifuge
Release time:
2013-11-28
Overview:
1. During holidays and evenings, you may leave only after performing the routine safety check on the centrifuge as the last step.
2. Only use high-quality centrifuge tubes; do not use centrifuge tubes that are aged, deformed, or cracked.
3. During the centrifugation process, operators must not leave the centrifuge room. If any abnormal situation occurs, operators must not turn off the power supply (POWER); instead, they should press STOP. Before pre-cooling, operators must complete the centrifuge usage log.
4. If the instrument malfunctions or components are damaged during use, promptly contact the technical staff.
Operating Precautions
1. The centrifuge must be reliably grounded; when the machine is not in use, please unplug the power cord.
2. It’s best not to run the centrifuge for more than 60 minutes at a time.
3. The centrifuge tubes must be filled with equal amounts of sample; never operate the rotor under unbalanced conditions.
4. Do not open the cover door while the machine is running or before the rotor has come to a complete stop, to avoid accidents.
5. Before use, check the rotor for any signs of damage, corrosion, or other defects. Also, inspect the centrifuge tubes for cracks, aging, and other issues. If you notice any doubts or concerns, immediately stop using the equipment and contact the manufacturer. Before starting the machine, be sure to tighten the clamping nut on the rotor securely to prevent the high-speed rotating rotor from flying off and causing an accident.
6. If you see “0.00” or any other number displayed during operation, and the machine fails to run, immediately turn off the power and unplug the machine. After waiting 10 seconds, plug it back in. Once the set speed is displayed, press the “Run” button, and the machine will resume normal operation.
7. The speed setting must not exceed the maximum speed to ensure safe operation of the machine.
8. If the specific gravity of the sample to be separated exceeds 1.2 g/cm³, the maximum rotational speed N must be corrected according to the following formula: N = NMAX × (1.2 / sample specific gravity)¹/², where NMAX is the rotor’s maximum allowable rotational speed.
9. To ensure safety and centrifugal performance, the instrument must be placed on a sturdy, level work surface; no items should be placed on the engineering plastic cover door. Samples must be arranged symmetrically, and before turning on the machine, make sure that all nuts have been securely tightened.
5. During precooling, the rotor lid can be placed either on the centrifuge’s platform or on the lab bench. Under no circumstances should it be left floating loosely on the rotor without being securely tightened, as an accidental start-up could cause the rotor lid to fly off and lead to an accident! No.
6. When the centrifuge is in the pre-cooling state, the centrifuge lid must be closed. After the centrifugation is complete, when removing the rotor, place it upside down on the experimental bench, wipe dry any remaining water inside the chamber, and leave the centrifuge lid open.
7. After tightening the rotor cover, be sure to use your fingers to check for any gaps between the rotor and the rotor cover. If you find a gap, loosen and retighten the cover until you confirm that there is no gap before starting the centrifuge.