Uses of laboratory medical centrifuges
Release time:
2013-11-01
Overview:
Medical centrifuges are widely used in medical research and hospital testing. Currently, the most commonly used centrifuge models in hospitals include high-speed centrifuges, blood centrifuges, blood-typing card centrifuges, urine sediment centrifuges, capillary centrifuges, cell centrifuges, and biopharmaceutical centrifuges—each with a wide variety of types and distinct functions. This article provides a brief introduction to the uses, functions, application types, as well as the proper usage and management and maintenance guidelines for medical centrifuges.
(1) Uses and Functions of Medical Centrifuges
1. Used to separate particles suspended in a solution.
1) Separate the formed elements from blood and concentrate cells or other formed elements from body fluids for analytical measurement.
2) Separate ligands bound to proteins or antibodies, as well as free ligands.
3) Separate the proteins that have already precipitated from the sample.
2. Separate two immiscible liquids of different densities.
1) If organic solvents are used to extract certain components from body fluids.
2) Separate lipid components from blood, such as isolating chylomicrons and various lipoproteins from plasma.
(2) Types of Applications for Medical Centrifuges
1. Horizontal rotor or bucket-type.
2. Fixed-angle rotor or angle rotor
3. High-speed centrifugal.
All three of the above categories include floor-standing and desktop models.
(3) Principle of Centrifugal Force
Its centrifugal force is expressed in terms of relative centrifugal force (RCF).
RCF = 1.118 × 10⁻⁵ × r × K²
In the formula: 1.118×10⁻⁵ is an experimental constant; r represents the radius (in cm) from the center of the rotation axis to the bottom surface of the centrifuge tube; and n denotes the rotational speed.
(4) Operation of Medical Centrifuges
To ensure the centrifuge operates properly, it is essential to use centrifuge tubes supplied by reputable manufacturers. Centrifuge tubes, rotor hubs, and rotor buckets must be balanced, with symmetrical placement at both ends of the rotor axis. During centrifugation, the maximum weight on each end must not exceed the limits specified by the manufacturer. When using high-speed centrifuges, operation must be performed by a dedicated operator, and a technician must double-check the rotational speed and the rotor being used.
(5) Management and Maintenance of Medical Centrifuges
1. If a centrifuge tube containing a sample or an empty tube breaks during centrifugation, immediately remove the glass shards and disinfect the centrifuge.
Disinfect the bucket or rotor and other centrifuge openings. In case of radioactive contamination, remove the contamination according to regulations and keep a record.
2. Wipe the inside and outside of the centrifuge with a damp cloth at least once a week.
3. Check the balance of the centrifuge and lubricate it once every 3 months.
4. Inspect the carbon brushes every 3 months and replace them promptly if they are worn out.
5. Check every 3 months whether the rotational speed matches the indicator reading.