Characteristics of Five Commonly Used Rotor Types for Centrifuges

Release time:

2012-09-20


Overview:

The rotational speed of a centrifuge rotor is related to the material and strength of the rotor. Generally, ultra-hard aluminum alloy (LC4), which has high strength and light weight, is used. For ultrahigh-speed centrifuges, titanium alloy (TiC4) is employed.

The rotor is the core component of a centrifuge used for separating samples. The number and variety of rotor specifications serve as an important indicator of the level of mastery in centrifuge manufacturing technology. Rotors can generally be categorized into the following five major types:

Fixed-Angle Rotor:

It is primarily used for separating particle samples whose sedimentation velocities differ significantly. The particles move only a short distance within the fan-shaped liquid layer; those that collide with the outer wall slide along the tube wall and settle at the bottom of the tube, thus enabling this rotor to quickly collect the precipitate. During the separation process, however, the presence of the tube wall induces vigorous convection within the centrifuge tube, which can adversely affect particles with identical sedimentation rates.

Horizontal rotor (Swing-Bucket Rotor):

It is primarily used for density-gradient centrifugation of samples. The particles travel long distances, and the positions of the components relative to the tube wall remain unchanged throughout and after centrifugation; thus, the centrifugation effect is excellent. However, since the particles in the centrifugal field are dispersed radially outward from the rotation rather than settling along parallel paths, they may collide with the outer wall and slide down along the tube wall, thereby inducing convection (though to a lesser extent than with fixed-angle rotors). At low speeds, vibrations during startup or shutdown can affect the separation efficiency. Swing-out rotors come in two types: open and closed. Generally, those with larger preparation capacities and rotational speeds below 10,000 RPM, where the centrifugal force field is less than 16,000×g, are designed as open-type rotors and are mainly used for the initial separation of samples. Rotors with smaller preparation capacities, higher rotational speeds above 10,000 RPM, and centrifugal force fields exceeding 16,000×g are typically designed as closed-type rotors to minimize the impact of air currents. These closed-type rotors are primarily used for the separation of organelles such as mitochondria, chloroplasts, and nuclei, as well as for density-gradient centrifugation.

Continuous-Flow Rotor:

It is primarily used for high-speed separation of smaller particulate matter in suspension media—for example, separating cells from culture media. During operation, the suspended sample flows into the centrifugal chamber from the center of the rotor at a certain velocity. Due to differences in the densities of the components in the solution, under the action of the centrifugal force field, heavier particles experience greater inertial centrifugal forces and settle more rapidly. When the flow rate increases (or decreases) to a specific value, the heavier particles will remain at the bottom along the outer edge of the centrifugal chamber, while the lighter particles will be carried away by the medium through the outlet. The separated components can be collected either by flushing them out of the centrifugal chamber or by retaining them within the rotor itself. This type of rotor employs centrifugal elutriation technology to separate whole cells or large subcellular particles. Under sterile and low-temperature conditions, the separated components maintain their activity and exhibit high recovery rates. If the direction of the liquid flow is opposite to that of the centrifugal force—that is, if it reduces the particle’s running speed—it effectively increases the particle’s travel distance, thereby enhancing the resolution capability.

Zonal-Rotor:

It is primarily used for high-capacity density gradient centrifugation. When the amount of sample required for an experiment is large and the swinging-bucket rotor is limited by its capacity, zone centrifugation is a better choice. The system mainly consists of a rotor body equipped with four blades and a sealing system. The rotor body and the four blades divide the centrifuge chamber into four fan-shaped compartments, preventing the solution from rotating freely within the chamber. Each blade is fitted with radial channels through which the solution flows from the center toward the outer wall of the rotor body. Care must be taken during sample loading and retrieval. Sample loading is typically performed while the rotor is spinning at 2,500 to 3,000 rpm. First, the lighter portion of the gradient solution is pumped into the rotor via a sealed pipeline. Due to inertial forces, this lighter gradient solution forms a uniform layer along the vertical direction on the outer wall of the rotor. Next, gradient solutions with progressively higher densities are pumped in one after another. As the density increases and more gradient solution is added, the lower-density gradient solution is pushed toward the center until the rotor is completely filled. Then, the sample to be separated—whose density is still lower than that of the lightest fraction in the gradient solution—is introduced through the central infusion tube of the sealing assembly. Subsequently, a less dense solution is injected to displace the sample near the center. Once sampling is complete, the sealing assembly is removed, and the rotor cover is securely closed. After centrifuging at a certain speed for a specified duration, the rotor speed is reduced to 2,500 to 3,000 rpm. The sealing assembly and infusion tube are then reinstalled, and denser liquids are pumped into the rotor, displacing the lighter solutions from the center outward in sequential order according to increasing density.

The zone-based rotor eliminates wall effects and interference caused by centrifuge tubes, thereby improving resolution.

Vertical-Tube-Rotor:

This method is primarily used for density-gradient centrifugation of samples over short periods. The swinging-bucket rotor enables particles to undergo long displacement distances and offers high resolution; after separation, the resulting bands are relatively broad, making it easy to recover the desired components. However, this method requires a longer centrifugation time. In contrast, with a vertical rotor, the particle displacement distance equals the diameter of the centrifuge tube. Before the rotor starts rotating, the density gradient solution inside the tube varies along the direction of gravity. Once rotation begins, the density of the solution gradually shifts to vary along the direction of centrifugal force (the horizontal direction). After centrifugation is complete, due to the influence of gravity, the density gradient reverts to varying vertically again, thus forming a vertical density gradient. Since the centrifuge tube is positioned vertically, the displacement distance of the solution particles is equal to the diameter of the tube. With shorter centrifugation times, the bands formed by sample components after centrifugation tend to be broader. However, because the centrifugation time is too short, particles may not achieve sufficient separation, leading to the loss of certain components. Moreover, given the large contact area between the bands and the gradient medium, diffusion tends to occur more readily. After the centrifuge is stopped, the density gradient realigns itself, causing the bands to mix easily. Therefore, in terms of banding effects, the vertical rotor is inferior to the swinging-bucket rotor.

The rotational speed of a centrifuge rotor is related to the material and strength of the rotor. Generally, ultra-hard aluminum alloy (LC4), which boasts high strength and light weight, is used. For ultrahigh-speed centrifuges, titanium alloy (TiC4) is employed. As a general rule, for a given centrifuge, rotors that are lighter in weight and have smaller capacities will operate at higher speeds, whereas those with greater weight and larger capacities will rotate at lower speeds. The maintenance and lifespan management of rotors will be discussed in subsequent chapters.