A Further Discussion on the Operating Guidelines for High-Speed Centrifuges in Laboratories

Release time:

2012-04-23


Overview:

When the centrifuge is in the pre-cooling state, the centrifuge lid must be closed. After the centrifugation is complete, once the rotor is removed, it should be inverted onto the lab bench, and any remaining water inside the chamber should be wiped dry. The centrifuge lid should remain open.
1. When the centrifuge is in the pre-cooling state, the centrifuge lid must be closed. After the centrifugation is complete, remove the rotor and place it upside down on the experimental bench. Wipe dry any remaining water inside the chamber, and leave the centrifuge lid open.
2. During precooling, the rotor lid can be placed either on the centrifuge’s platform or on the lab bench. Under no circumstances should it be left floating loosely on the rotor without being securely tightened, as an accidental start-up could cause the rotor lid to fly off, leading to an accident!
3. After tightening the rotor cover, be sure to use your fingers to check for any gaps between the rotor and the rotor cover. If there is a gap, loosen and retighten the cover until you confirm that there are no gaps before starting the centrifuge.
4. During the centrifugation process, operators must not leave the centrifuge room. If any abnormal situation occurs, operators must not turn off the power supply (POWER); instead, they should press the STOP button. Before pre-cooling, operators must complete the centrifuge usage log.
5. Do not use substandard or counterfeit centrifuge tubes, nor use centrifuge tubes that are aged, deformed, or cracked.
6. During holidays and in the evening, you may leave only after performing the routine safety check following the last use of the centrifuge.
7. In the event of a machine malfunction or component damage during instrument use, please contact the manufacturer promptly—phone: 021-68113426.