Daily Maintenance and Care for Benchtop Centrifuges
Release time:
2012-04-11
Overview:
When using a desktop centrifuge in our daily routine, we should first disconnect the power supply and release the centrifuge’s brake before starting it. We can then manually rotate the rotor to check whether there is any binding or jamming. Check other parts for looseness and any abnormal conditions.
Connect the power supply and start the machine by turning the knob clockwise (typically, it takes about 40 to 60 seconds from a standstill to normal operation).
Typically, each piece of equipment must be run empty for about 3 hours after it arrives at the factory; it can begin operation once no abnormalities are detected.
Materials should be placed as evenly as possible.
It must be operated by a dedicated person, and the capacity must not exceed the rated capacity.
It is strictly prohibited to operate the machine at speeds exceeding its rated capacity, as this may shorten the machine's service life.
After the machine is started, if any abnormal conditions occur, it must be stopped for inspection and, if necessary, disassembled, cleaned, and repaired.
When operating, a desktop centrifuge spins at high speed; therefore, never touch its rotor with your body to prevent accidents.
The mesh count of the filter cloth should be determined according to the size of solid particles in the material being separated; otherwise, the separation efficiency will be affected. In addition, when installing the filter cloth, the sealing ring of the filter cloth must be properly embedded into the sealing groove of the rotating drum to prevent material from leaking out.
To ensure the centrifuge operates properly, please lubricate and maintain the rotating components every six months. At the same time, check the lubrication condition at the bearings for any signs of wear; inspect the components in the braking system for wear, and replace any severely worn parts; and verify whether there are any oil leaks from the bearing covers.
After the machine has been used, it should be thoroughly cleaned to keep it tidy.
Do not use non-corrosion-resistant centrifuges for separating highly corrosive materials. In addition, operate strictly in accordance with the equipment’s requirements and specifications; never use non-explosion-proof centrifuges in flammable or explosive environments.
Previous: