Introduction to New Technologies from Shanghai Medical Centrifuge Manufacturers
Release time:
2012-02-09
Overview:
1.1 Temperature Cooling
When the rotor spins at high speed, air friction generates heat, causing the rotor's temperature to rise and, consequently, the temperature of the sample in the tube to increase as well. Since biological samples are highly sensitive to temperature, centrifugation experiments generally require that the temperature be maintained at 4℃. Without intelligent technology controlled by a computer, it would be impossible to achieve an accuracy of ±1℃. Therefore, refrigeration units are used to cool the centrifuge chamber, thereby cooling both the rotor and the sample. However, measuring the actual temperature of a rotating rotor is extremely challenging. Both domestically and internationally, temperature sensors are typically placed near the bottom of the centrifuge chamber, close to the rotor, to indirectly measure the rotor’s temperature. Here, T_sample = T_chamber + T_compensation. The compensation temperature (T_compensation) depends on the rotor’s type, its rotational speed, and the duration of operation.
1.2 Direct drive with brushless motor
Centrifuge It is driven by a motor that rotates the rotor at high speed. In the past, motors were DC motors equipped with carbon brushes. During operation of the centrifuge, these carbon brushes wore down, producing sparks, noise, and even vibrations. Moreover, carbon brushes have a limited lifespan, so they had to be replaced periodically before the machine could resume operation. Even more serious was the pollution caused by the wear of carbon brushes, which generated carbon dust. This not only contaminated the centrifuge itself but also polluted the surrounding environment—a type of contamination that is entirely unsuitable for the healthcare industry.
1.3 Display digital technology
A typical manifestation of analog technology is the use of knobs to select operational parameters—such as rotational speed, temperature, and time—and displaying the data via pointers on a dial. Its drawbacks include significant human-induced variability in parameter selection and data reading, leading to poor control accuracy. Moreover, this process must be repeated every time, resulting in poor repeatability. Although some centrifuges feature digital displays (e.g., using knobs to set values while the digits display the readings), despite improvements in readability, the underlying principle for value acquisition remains analog technology; thus, these still fall under the category of analog technology. In contrast, digital displays typically exhibit user-friendly interfaces, keyboard-based operation, digitalized displays, and fully computer-controlled, programmable functions. At their core lies intelligent control, which is realized entirely by computers. Centrifuge All parameters required for the operation—such as rotational speed, temperature, duration, acceleration/deceleration rates, and others—are entered via keyboard input and displayed digitally. As a result, selecting operational parameters and reading the displayed values involves minimal error. Moreover, since the operation is programmable, a set of operational parameters can be assigned a specific number and stored for future retrieval and use.